Commissioning
PurCon
62
3.5
Installing the PurCon air outlet (option)
We recommend to install the PurCon air outlet if the drain pipe is not
entirely pressure-free. By using the air outlet, no overpressure can
develop in the system which prevents the retentate from running off.
The PurCon air outlet is mounted on the filler nozzle instead of the cap.
Fig. 8
PurCon air outlet
Installation
PurCon
3
Disable the feed transport.
4
Switch off the power switch of the PurCon.
5
Unscrew the cap of the filler nozzle on the top of the housing.
6
Put the PurCon air outlet on the filler nozzle as shown in Fig. 8
and screw it tight by hand.
7
To avoid obnoxious odors, plug a suitable DN 30 tube on the
tube nozzle and lead the other end of the tube into the open or
into a discharge.