PurCon
What to do if ...
83
6
What to do if ...
Permeate is not being
fed
Warning
Never let any water get into the inside of the enclosure!
1
Is the permeate pump turning?
If necessary, check the power supply of the PurCon.
If the fault is not remedied, continue with step 2.
2
Are the pump tubings properly mounted and
functioning?
Replace pump tubings if necessary.
If the fault is not remedied, continue with step 3.
3
Is the submersible pump working?
If necessary, check the power supply and the functioning of the
submersible pump.
If the fault is not remedied, continue with step 4.
4
Clean the sample preparation instrument according to section
C
LEANING
THE
SAMPLE
PREPARATION
INSTRUMENT
. If the fault is
not remedied by thorough cleaning, the filter plates have to be
replaced (see chapter 6.1 R
EPLACING
THE
FILTER
PLATES
).