What to do if ...
PurCon
88
LSet/PC
When exchanging the tubes, observe the following points:
l
Make sure the original tubes and the spare tubes have the same
length. The correct lengths are given in Fig. 23.
l
Secure all tube connections with cable ties.
l
Run the air tube in a loop so that its highest point is above the drain
pipes of the filter unit (see Fig. 23). Otherwise, liquid may flow back
into the air compressor when the PurCon is switched off.
Fig. 23
Air carrying tubes of the PurCon