Multi 9310
Dissolved oxygen
ba75937e03
12/2011
55
Calibration
evaluation
After the calibration, the meter automatically evaluates the current status of the
calibration. The evaluation appears on the display and in the calibration record.
FDO
®
925
calibration
evaluation
Calibration record
(USB output)
Menu item
Setting/
function
Description
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration records.
Further options:
z
Scroll through the calibration
records with
<
S
S
S
S
><
T
T
T
T
>
.
z
Output the displayed calibra-
tion record to the interface with
<F2>
/
[USB output]
.
z
Output all calibration records
to the interface with
<F2__>
[USB output]
.
z
Quit the display with
<F1>
/
[Back] or
<ENTER>
.
z
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to
RS232/USB
-
Outputs the calibration data stor-
age to the interface
Display
Calibration record
Relative slope
+++
S = 0.94 ... 1.06
++
S = 0.92 ... 0.94 or
S = 1.06 ... 1.08
+
S = 0.90 ... 0.92 or
S = 1.08 ... 1.10
Error
Error
S < 0.90 or
S > 1.10
Error elimination (see section 14 W
CALIBRATION Ox
Calibration date
01.07.2011 07:43:33
FDO 925
Ser. no. 10146858
SC-FDO 925
10158765
Relative slope
0.98
Sensor
+++