Data storage
Multi 9310
76
ba75937e03
12/2011
10.2 Automatic data storage at intervals
The data storage interval (
Interval
) determines the time interval between auto-
matic data storage processes. Each data storage process transmits the current
dataset to the interface at the same time.
Configuring the
automatic data
storage function
Settings
You can configure the automatic data storage function with the following set-
tings:
1.
Press the
<STO__>
key.
The menu for automatic data storage appears.
1 Adjusted entire data storage duration
2 Max. available data storage duration
3 Graphical display of the data storage usage
ID number
1
Interval 30
s
Duration 180
min
continue
0d03h00min
0
1d17h33min
Automatic data storage
1
2
3
01.07.2011
08:00
Back
Menu item
Possible setting
Description
ID number
1 ... 10000
ID number for the dataset series.
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min, 10 min,
15 min, 30 min,
60 min
Storage interval.
The lower limit of the storage interval
can be restricted by the number of free
storage locations.
The upper limit is restricted by the stor-
age duration.
Duration
1 min ... x min
Storage duration.
Specifies after which time the auto-
matic data storage should be termi-
nated.
The lower limit of the storage duration
is restricted by the storage interval.
The upper limit is restricted by the
number of free storage locations.
Summary of Contents for Multi 9310