Multi 9310
What to do if...
ba75937e03
12/2011
91
Display
Instrument does not
react to keystroke
You want to know
which software
version is in the
meter or IDS sensor
Cause
Remedy
– Batteries almost empty
– Replace the batteries (see section
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<ENTER>
and
<On/Off>
key simultaneously
Cause
Remedy
– E. g., a question by the service
department
– Switch on the meter.
– Open the menu,
<ENTER__>
/
Stor-
age & config
/
System
/
Service infor-
mation
. The instrument data are
displayed.
or
– Connect the sensor.
Press softkey [
<F1>
Info
<F1>
] /
[
More
] The sensor data are displayed