background image

Operation

IQ S

ENSOR

N

ET

 MIQ/T2020 PLUS

4 - 14

Safety copy before

changing the system

configuration

If you want to carry out changes of the system configuration that may 
have to be undone again (for example if you want to test certain 
functions for a short time), you can perform a manual saving process 
beforehand. When doing so the timer for automatic saving is reset, i. e.  
the safety copy is available for 60 minutes. Then it is overwritten with 
the current configuration.

4.3.3

Retransferring the system configuration

The system configuration saved by the MIQ/T2020 PLUS can be 
retransferred to the main controller. Thus, the original system 
configuration can be quickly restored after exchanging a defective 
controller. Precondition for this is that both controllers are of the same 
type and the software version of the replacement controller is equal or 
higher than that of the original controller.

Note
With controller versions lower than 2.70, the following settings cannot 
be retransferred:

l

Password

l

Locking open or closed

l

Language

l

Height of the location (with systems without automatic air pressure 
compensation only).

The settings of the main controller are retained here.

Retransfer after MIQ/

T2020 PLUS controller

operation

After being initialized as the regular terminal, the MIQ/T2020 PLUS 
checks whether the system configuration in the main controller differs 
from that in the MIQ/T2020 PLUS (date and time are not taken into 
account). If the configurations differ, a window appears on the display 
containing a corresponding note and the offer to retransfer the system 
configuration to the main controller.

Note
The timer for the regular automatic saving process carried out by the 
MIQ/T2020 PLUS only starts when you have confirmed or rejected the 
command for retransferring the system configuration. From this point of 
time, it takes 60 minutes for the first automatic saving process to start 
(there is the additional option to save manually). If you reject the 
command to retransfer the system configuration and continue to work 
with the system configuration of the main controller, the system 
configuration stored up to this time is overwritten in the MIQ/
T2020 PLUS after 60 minutes and cannot be restored anymore.

Summary of Contents for MIQ/T2020 PLUS

Page 1: ...ba64126e01 03 2004 MIQ T2020 PLUS Operating Manual C M S ESC OK MIQ T2020 PLUS IQ SENSOR NET Terminal with redundant controller function...

Page 2: ...cannot guarantee that there are absolutely no errors in this manual Therefore we are sure you will understand that we cannot accept any legal claims resulting from the data figures or descriptions Not...

Page 3: ...on 3 3 4 Operation 4 1 4 1 Introduction 4 1 4 1 1 Overview of the operating elements 4 2 4 1 2 Display 4 3 4 1 3 Keys 4 5 4 1 4 Rotary switch 4 5 4 2 General operating principles 4 6 4 2 1 Navigating...

Page 4: ...Contents IQ SENSOR NET MIQ T2020 PLUS 0 2...

Page 5: ...lease file this component operating manual into the ring binder of the system operating manual 1 2 Terminal function of the MIQ T2020 PLUS The MIQ T2020 PLUS terminal is an extension component for the...

Page 6: ...at should be displayed at this measurement location This function local identity function can be used e g to quickly find sensors for calibration on site The information is transmitted by the two uppe...

Page 7: ...hout restrictions Restrictions in the MIQ T2020 PLUS controller operation In the MIQ T2020 PLUS controller operation the following functions cannot be supported l Editing the system configuration Apar...

Page 8: ...Overview IQ SENSOR NET MIQ T2020 PLUS 1 4...

Page 9: ...th the system operating manual and all other component operating manuals in the vicinity of the IQ SENSOR NET system General safety instructions The following safety labels in the individual chapters...

Page 10: ...secure technical condition Function and operational safety The failure free function and operational safety of the MIQ T2020 PLUS is only guaranteed if the generally applicable safety measures and th...

Page 11: ...SENSOR NET main controller with the software version 2 04 or higher Full functionality is available when using a main controller with the software version 2 70 or higher Using older main controller v...

Page 12: ...ing the terminal Removing the terminal To remove the terminal press the locking lever and pull the terminal forwards 1 Pull the cover off the contacts pos 1 in Fig 3 1 on the lid of the MIQ module 2 P...

Page 13: ...s soon as contact is established to the IQ SENSOR NET In the first phase the background illumination of the display switches on and the terminal is initialized This process may last a few seconds The...

Page 14: ...Installation IQ SENSOR NET MIQ T2020 PLUS 3 4...

Page 15: ...ibration l Perform system and terminal settings l Display messages Note To guarantee data safety within the IQ SENSOR NET access to the following functions is only possible from one terminal at the sa...

Page 16: ...display for the clear presentation of current measured values the graph of measured values status displays and message texts The IQ SENSOR NET system is operated using the 5 keys m c s e g and the ro...

Page 17: ...a open settings released System settings can be changed z closed settings locked All system settings can only be read 6 Error symbol h If the error symbol flashes a new or unacknowledged error message...

Page 18: ...alue display The measured value display contains the following information for each sensor differential sensor Fig 4 3 Display measured value display 1 Consecutive numbering of the measured values 2 M...

Page 19: ...ith sensor interrupted OFL Measuring range undercut or exceeded overflow Clean The time controlled sensor cleaning function is active see MIQ CR3 operating manual Display flashes Sensor in maintenance...

Page 20: ...ields Select a setting in selection fields Select a character in text entry fields l Confirm a selection with the g key l Break off an action and change to the next higher level with the e key l Start...

Page 21: ...wn here 2 Select a menu item e g System settings with d Move the highlighting reverse video display black background in the list of menu items by turning the rotary switch d 3 Confirm the menu item e...

Page 22: ...Open the Settings menu with s 2 Select the Edit list of sensors menu item with d 3 Confirm the menu item e g Edit list of sensors with g The Edit list of sensors display opens A column is highlighted...

Page 23: ...Edit list of sensors Note The following letters numerals and special characters can be entered AaBb Zz0 9 _ Fig 4 7 120 Edit list of sensors 8 Select a letter or numeral with d 9 Confirm the letters...

Page 24: ...m on the following page shows the functionality of the MIQ T2020 PLUS controller operation 10 l Add a new character Select the character to be added with d and confirm with g or l Delete the last char...

Page 25: ...Regular terminal operation Saving of the system configuration every 60 min Saving of system configuration every 60 min from now on If settings are different Retransfer of system configuration Confirm...

Page 26: ...rocess takes place after 60 min A message appears on the display during the saving process If required the running saving process can be canceled with the e or m key l The MIQ T2020 PLUS continuously...

Page 27: ...iguration to the main controller see section 4 3 3 Note With controller versions 2 66 or higher the actual system configurations are compared by means of a checksum This is not possible with older con...

Page 28: ...pen or closed l Language l Height of the location with systems without automatic air pressure compensation only The settings of the main controller are retained here Retransfer after MIQ T2020 PLUS co...

Page 29: ...9 100 Settings Service 1 Open the Settings menu with s 2 Select the Service menu item with d and press g 3 Using d select the menu item Retransfer configuration and press g The process is started 4 Wa...

Page 30: ...Operation IQ SENSOR NET MIQ T2020 PLUS 4 16...

Page 31: ...MIQ CR3 appear in the standard language English To activate the selected system language for this component a software update of the component is required Please contact WTW 1 Open the Settings menu w...

Page 32: ...nd the instrument designation alternately appear in the top left corner of the display display changes in cycles of 5 s 15 characters are available for the name 1 Open the Settings menu with s 2 Selec...

Page 33: ...adjusted Display contrast The contrast level of the display can be continuously adjusted 5 Select the Illumin brightness menu item with d and confirm with g 6 Select a brightness level and confirm wit...

Page 34: ...Settings Setup IQ SENSOR NET MIQ T2020 PLUS 5 4...

Page 35: ...1 Maintenance The MIQ T2020 PLUS requires no special maintenance The general maintenance of IQ SENSOR NET components is described in the IQ SENSOR NET system operating manual 6 2 Cleaning The cleaning...

Page 36: ...Maintenance and cleaning IQ SENSOR NET MIQ T2020 PLUS 6 2...

Page 37: ...terminal dimensions in mm Mechanical construction Side view Front view 42 0 37 0 210 0 170 0 137 185 Stacked mounting 37 0 48 0 Housing material ASA Acrylonitrile Styrene Acrylesterpolymer Weight App...

Page 38: ...pending on the number of terminal components already installed A total of up to three terminal components can be used in the IQ SENSOR NET Electrical data Supply voltage Max 24 VDC via the IQ SENSOR N...

Reviews: