Operation
IQ S
ENSOR
N
ET
MIQ/T2020 PLUS
4 - 12
How it works in detail:
Normal operation
l
In normal operation, the main controller carries out the regular
controller functions.
l
The MIQ/T2020 PLUS works as a regular terminal.
l
The MIQ/T2020 PLUS saves the system configuration:
– Automatic saving processes take place 60 minutes after
initialization and then regularly every 60 min.
– Additionally to automatic saving, the saving can also be started
manually at any time (see section 4.3.2). When doing so the timer
for automatic saving is reset.
– Automatic saving only takes place in the measuring mode
(normal measured value display). If the MIQ/T2020 PLUS is in a
different mode, the next saving process takes place as soon as it
is switched to the measuring mode.
– If a saving process fails, e.g. because another terminal is in the
setting or calibration mode, a new saving process is started after
30 s. After three failed trials the next automatic saving process
takes place after 60 min.
– A message appears on the display during the saving process. If
required, the running saving process can be canceled with the
e
or
m
key.
l
The MIQ/T2020 PLUS continuously receives information on the
status of the main controller.
Event: Failure of the
main controller
(MIQ/T2020 PLUS
controller operation)
l
If the MIQ/T2020 PLUS does not receive a valid telegram from the
main controller for a period of 2 minutes it takes over the controller
operation. The MIQ/T2020 PLUS is reinitialized.
l
A log book message is generated when the controller operation is
taken over.
l
During the MIQ/T2020 PLUS controller operation
– an "A" (=active) appears behind the instrument designation in the
top left corner of the display.
– no changes of the IQ S
ENSOR
N
ET
configuration can be made.
Only the date and time can be changed. All other settings are
locked (
z
).
– neither sensors nor input or output modules can be added.
– no data logger is available. Accordingly, the data transmission
function cannot be carried out.
– the MIQ/T2020 PLUS keeps a separate log book. The log book
is empty when the controller operation is taken over. The topmost
Summary of Contents for MIQ/T2020 PLUS
Page 4: ...Contents IQ SENSOR NET MIQ T2020 PLUS 0 2...
Page 8: ...Overview IQ SENSOR NET MIQ T2020 PLUS 1 4...
Page 14: ...Installation IQ SENSOR NET MIQ T2020 PLUS 3 4...
Page 30: ...Operation IQ SENSOR NET MIQ T2020 PLUS 4 16...
Page 34: ...Settings Setup IQ SENSOR NET MIQ T2020 PLUS 5 4...
Page 36: ...Maintenance and cleaning IQ SENSOR NET MIQ T2020 PLUS 6 2...