
pH 170/
pH 296
Checking Mode
166
Recorder output 1
A current of 0.1 mA is set in the
tSt
rEC1
submenu after pressing the ENTER key:
−
Press C to set the current to 20.0 mA
−
Press M to set the current to 0.1 mA
−
Press UP to increase the current
(max. 20.5 mA)
−
Press DOWN to decrease the current
(min. 0.0 mA)
−
Press the ENTER key again to continue with
recorder output 2
Recorder output 2
A current of 0.1 mA is set in the
tSt
rEC2
submenu after pressing the ENTER key:
−
Press C to set the current to 20.0 mA
−
Press M to set the current to 0.1 mA
−
Press UP to increase the current
(max. 20.5 mA)
−
Press DOWN to decrease the current
(min. 0.0 mA)
−
Press the ENTER key again to continue with
relay test.