Parameterization
pH 170/
pH 296
141
REC 1 and REC 2 recorder outputs
(REC 2 only in T option)
This submenu enables the setting of:
•
Current range
−
0 to 20 mA
−
4 to 20 mA
•
Attenuation [dI/dt] (dI = current change; dt = time change)
−
0.1 mA/s
−
1.0 mA/s
−
5.0 mA/s
−
20.0 mA/s
(= delivery state)
Note:
Attenuation = changing speed of the recorder flow [mA/sec]
on a sudden change of the input signal.
Setting:
−
Press the ENTER key when the
Pr1
submenu appears:
−
Use the UP / DOWN keys to select the current
range.
−
Press the ENTER key to confirm the input.
−
Use the UP / DOWN keys to select the
attenuation.
−
Press the ENTER key to confirm the input.
−
Press the ENTER key when the
Pr2
submenu appears:
−
Set recorder output 2 in the same way as
recorder output 1.
−
Press the ENTER key to confirm the input.