Settings
Cond 7310
30
ba75933e04
06/2013
6.1.2 Calibration interval
The calibration evaluation is displayed as a sensor symbol.
The sensor symbol flashes after the adjusted calibration interval has expired. It
is still possible to measure.
Setting the calibra-
tion interval
The calibration interval is set to 150 days in the factory.
You can change the interval (1 ... 999 days):
Measurement
/
Temp. comp.
(TC)
/
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for the
measured parameters,
ϰ
and
ρ
.
Measurement
/
TDS factor
0.40 ...
1.00
Factor for TDS value
Measurement
/
Stability control
On
Off
Switches on or off the automatic stabil-
ity control during measurement (see
section Stability control (AutoRead) &
HOLD function, page 23 )
Measurement
/
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees Fahren-
heit.
All temperature values are displayed
with the selected unit.
Reset
-
Resets all sensor settings to the deliv-
ery condition (see section 6.3.1 R
To ensure the high measuring accuracy of the meter, calibrate after
the calibration interval has expired.
Menu item
Possible
setting
Description
1.
Open the menu for measurement settings with [Menu]
<F1>
.
2.
In the
Calibration
/
Calibration interval
menu, set the calibration interval
with
<
><
>
.
3.
Confirm the setting with
<ENTER>
.
4.
Quit the menu with
<M>
.