
Cond 7310
Settings
ba75933e04
06/2013
29
Calibration
/
Calibration inter-
val
1 ...
150
...
999 d
Calibration interval
for the measuring
cell (in days).
The meter reminds you to calibrate
regularly by the flashing sensor symbol
in the measuring screen.
Measurement
/
Measuring cell
/
Type
Measuring cell
used
Cal
Measuring cells whose cell constant is
determined by calibration in the KCL
control standard solution.
Calibration ranges:
0.450 ... 0.500 cm
-1
and
0.800 ... 0.880 cm
-1
The currently valid cell constant is dis-
played in the status line.
LR325/01
LR 325/01 measuring cell,
nominal cell constant 0.100 cm
-1
.
The cell constant can be adjusted in
the range 0.090 ... 0.110 cm
-1
.
LR325/001
LR 325/001 measuring cell,
nominal cell constant 0.010 cm
-1
.
The cell constant is permanently set.
man
Any measuring cells with freely adjust-
able cell constants in the range 0.250
... 25.000 cm
-1
.
Measurement
/
Measuring cell
/
Cell const. man
0.250 ...
0.475
25.000 cm
-1
Display and setting option of the cell
constant of any measuring cells (man).
Measurement
/
Measuring cell
/
Cell const.
LR325/01
0.090 ...
0.100
...
0.110 cm
-1
Display and setting option of the cell
constant of the LR 325/01 measuring
cell.
Measurement
/
Temp. comp.
(TC)
/
Method
nLF
Lin
Off
Procedure for temperature compensa-
tion (see section 6.1.3 T
This setting is only available for the
measured parameters,
ϰ
and
ρ
.
Measurement
/
Temp. comp.
(TC)
/
Linear
coeff.
0.000 ...
2.000
...
10.000 %/K
Coefficient of the linear temperature
compensation.
This setting is only available when the
linear temperature compensation is
set.
Menu item
Possible
setting
Description
Summary of Contents for Cond 7310
Page 58: ...Index Cond 7310 58 ba75933e04 06 2013...
Page 59: ......