Cond 3310 IDS
Settings
ba77051e04
07/2017
35
6.3
Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
6.3.1 Resetting the measurement settings
Conductivity
The following settings for conductivity measurements are reset to the default
settings with the
Reset
function:
The sensor settings are reset under the
Reset
menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
<ENTER>
key.
6.3.2 Resetting the system settings
The following system settings can be reset to the default condition:
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Calibration interval
150 d
Measured parameter
ϰ
Cell constant (c)
Depending on the connected measur-
ing cell:
0.475 cm
-1
(calibrated)
0.475 cm
-1
(adjusted)
0.100 cm
-1
Temperature compensation
nLF
Reference temperature
25 °C
Temperature coefficient (TC) of
the linear temperature compensa-
tion
2.000 %/K
TDS multiplier
1,00
Setting
Default settings
Language
English
Audio signal
on
Baud rate
4800 Baud
Output format
ASCII