Settings
Cond 3310 IDS
32
ba77051e04
07/2017
6.2
Sensor-independent settings
6.2.1
System
To open the
Storage & config
menu, press the
<ENTER_>
key in the measured
value display. After completing the settings, switch to the measured value dis-
play with
<M>
.
Default settings are printed in
bold
.
Temp. comp. (TC)
/
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for the
measured parameters, conductivity
(
ϰ
) and resistivity (
ρ
).
Multiplier for TDS
0,40 ...
1,00
Multiplier for TDS value
Reset
-
Resets all sensor settings to the
delivery condition (see section
6.3.1 R
Menu item
Possible
setting
Explanation
Menu item
Possible setting
Explanation
System
/
General
/
Lan-
guage
Deutsch
English
(more)
Selects the menu language
System
/
General
/
Audio
signal
on
off
Switches on/off the beep on keystroke
System
/
General
/
Illumi-
nation
Auto
on
off
Switches the display illumination on/off
System
/
General
/
Con-
trast
0 ...
50
... 100 %
Changes the display contrast
System
/
General
/
Shutoff
time
10 min ...
1h
...
24 h
Adjusts the switch-off time
System
/
General
/
Tem-
perature unit
°C
°F
Temperature unit,
degrees Celsius or degrees Fahrenheit.
All temperature values are displayed with the
selected unit.
System
/
General
/
Stabil-
ity control
on
off
Switches on or off the automatic stability control
during measurement (see section 6.2.3 A
System
/
Interface
/
Baud
rate
1200, 2400,
4800
, 9600,
19200
Baud rate of the USB Device interface
Summary of Contents for Cond 3310 IDS