47 | UM-PP-EN-03
6.5.4
Remove Handrails
Side Panels
Loosen the screws on both side panels (3 per side) with a Phillips screwdriver.
On PPS Bari-Mill models, the cutout for the railing is covered with an additional panel. Loosen the
fixing screws on the panels (2 per side). See the following figure for the placement of side panel fixing
screws and note that there are fixing screws on both sides of the device.
Fig. 22 Side panel fixing screws
1.
PPS Bari-Mill side panel fixing screws
2.
PPS Med / Ortho side panel fixing screws
Electronics Covering
The drive and control electronics are located on the right side of the device as seen from walking
direction.
To remove the metal cover (shield) loosen the 4 fixing screws.
Fig. 23 Remove electronics shielding panel
1
2