2.3 Line Voltage Switch Does Not Operate Attached Device
Chapter 2: Troubleshooting Procedures
2.3 Line Voltage Switch Does Not Operate Attached Device
SHOCK HAZARD:
Use care when checking live voltages. Do not touch live terminals,
wires, and ground. Failure to use caution could result in personal injury.
1. All other Integris® 2001 Headwall System electrical systems are
operational.
Yes
No
↓
→
Restore electrical service from the building to the Integris®
2001 Headwall System.
2. If so equipped, the applicable Integris® Headwall 2001 Headwall System
circuit breaker is in the ON position.
Yes
No
↓
→
Reset the applicable circuit breaker to the ON position.
3. The line voltage switch is operative.
Yes
No
↓
→
4. Go to “Final Actions” on page 2-5.
5. Perform the following:
a. Gain access to the line voltage switch.
b. Place the switch in the OFF position.
c. Measure the voltage across the switch terminals.
A voltage of 109 to 125V AC is present.
Yes
No
↓
→
6. Perform the following:
a. Place the applicable circuit breaker in the OFF position.
b. Place the line voltage switch in the ON position.
c. Check for continuity across the switch terminals.
Continuity is present.
Yes
No
↓
→
Replace the line voltage switch. See “Line Voltage Switch” on
page 4-9.
Page 2 - 10
Integris® 2001 Headwall System Service Manual (man148ra)
Summary of Contents for Integris 2001
Page 4: ...Revisions NOTES Page iv Integris 2001 Headwall System Service Manual man148ra...
Page 12: ...Table of Contents NOTES Page xii Integris 2001 Headwall System Service Manual man148ra...
Page 14: ...Chapter 1 Introduction NOTES Page 1 2 Integris 2001 Headwall System Service Manual man148ra...
Page 288: ...Chapter 7 Accessories NOTES Page 7 2 Integris 2001 Headwall System Service Manual man148ra...