6
Power supply
The power supply is provided by means of a rechargeable battery pack with three integrated, re-
chargeable batteries (Type: see Technical data). The service life of the rechargeable battery pack
is dependent on the ambient conditions and the number of charge cycles.
If required, the rechargeable battery pack can be replaced. Only genuine WITT spare parts should
be used here, a new rechargeable battery pack can be ordered from WITT-Gasetechnik
.
The charge status of the rechargeable batteries is denoted by a battery symbol in the top right cor-
ner of the display, a dark battery indicates: fully charged. The Oxybaby
®
switches off when the
rechargeable battery is spent.
If the rechargeable battery charge is no longer sufficient to perform a reliable measurement,
a warning message is shown on the display. In such an instance do not perform any more
analyses, as the measurement accuracy may be impaired.
You can easily switch between rechargeable battery mode and power supply unit mode
when operating it.
6.1
Charging rechargeable batteries
We advise you to charge the rechargeable batteries for the Oxybaby
®
at the very latest when "bat-
tery low" is shown on the display.
Connect the Oxybaby
®
to the power supply unit / charge as follows and charge the batteries:
fold down lug on right side.
insert combined USB charging cable / power supply unit cable into USB
connection on the Oxybaby
®
.
insert the plug of USB charging cable / power supply unit cable into socket
on power supply (see Technical data) and charge the rechargeable batter-
ies.
The battery symbol at the top right of the display flashes on and off as
long as the rechargeable batteries are being charged.
Do not perform any measurements during the charging process, as the
measuring accuracy may be slightly impaired.
To ensure that the calibration accuracy is not impaired:
do not calibrate the device until 1 hour after the rechargeable batteries
have been charged.
6.2
Power supply unit operation
Use the USB power supply unit / charger as supplied only!
Otherwise the OXYBABY
®
may not function correctly. WITT-Gasetechnik will not be held
liable for any damage caused by external power supply units or chargers.
The USB connection is disabled when the power supply unit / charger is being used.
Fold down lug on right side.
Insert combined USB charging cable / power supply unit cable into USB connection on the
OXYBABY
®
.
Insert the plug of USB charging cable / power supply unit cable into socket on power supply
(see Technical data).
OXYBABY
®
switches on and is now ready to perform measurements
You can change into the rechargeable battery mode at any given time.