Proceed in accordance with the result of the measurements mentioned above:
Result of measurements.
What is to be done?
No difference in concentration levels
between measurement of a packaging and the measurement of
zero gas (without needle and filter).
Send device to WITT after-
sales service
Difference in concentration level
between the measurement of packaging and the measurement
of zero gas.
Replace filter (see below) /
check needle (see below)
Replace filter or needle:
Fig. 2
filter (1)
needle cover (3)
needle (2)
Unscrew needle cover
(3)
in counterclockwise direction.
Replace filter
(1)
.
Check needle
(2)
for flow rate:
pierce needle
(2)
with included fine wire. If this is not possible, replace needle.
If the fault persists, send the device to WITT after-sales service.
Caution:
never use the OXYBABY
®
without a filter!