7.3.2
Measurement in Permanent mode
Change the operating mode to “permanent” if required (see Chapter 8.4).
Select any product, user or line (production line) (see Chapter 8)
Stick foam pads onto the packaging that you wish to measure.
This seals the pierced position.
Open the needle cover (see 7.2)
Pierce the packaging through the foam pad.
Caution!
Do not jab into the fluid or the packaged product. Otherwise the OXYBABY will be soiled,
the needle or hose may clog up or the sensor may be damaged.
Start measurement with
.
End measurement with
.
Close the needle cover (see 7.2)
The last current measurement is shown on the display.
During the permanent measurement, the current measured value is automatically stored
every 10 seconds along with the date and time.
7.4 Switching
off
When in
rechargeable battery mode
:
(i.e. operation without USB power supply unit and without USB cable at the USB interface)
Press
key for approx. 2 seconds and hold.
The device is now automatically purged with air and it then switches off.
The OXYBABY
®
is also equipped with an automatic switch-off function.
It switches off automatically if no commands are received for approx. 2 minutes (e.g. key ac-
tuations or measurements).
If the power supply is connected through the
USB interface
(through plugged-in USB power supply
unit or a USB cable).
Disconnect USB power supply unit or USB cable.
Continue as described above in rechargeable battery mode.