background image

 

INSTRUCTION MANUAL 

NI-492WE 

Rev.4  05/18 

 

4 of 7 

9.2.2 TEMPERATURE SWITCHES CLASS SAMA II C 

Difference  in  height  between  bulb  and  instrument  enclosure 
when installed 

must not exceed

 two meters. 

With a difference of two meters the set point value fixed during 
the  calibration  may  be  affected  by  a  maximum  error  of  1.5°C. 
This error depending on the normal reference temperature, the 
working  temperature  and  the  set  point.  For  further  clarification 
request 

IS-TC.401E

9.3 

BULB AND CAPILLARY 

With reference  to  figure  5 unscrew  the  fitting  (3)  from  the  seal 
press (2) and slide it off from the bulb (5).  

Fig. 5 – Mounting of bulb 

 

1 -  Armored capillary 
2 -  Stuffing nut (SW 

12) 

3 -  Rotating fitting (SW 

22) 

4 -  Thermowell 
5 -  Bulb 
6 -  Stainless steel 

washer 

7 -  PTFE washer 

 

Mount the fitting (3) on the thermowell (4) and tighten it using the 
appropriate key. Rid the capillary from the armored by pulling the 
bulb (5) with one hand while holding the stuffing nut with the other 
(2). 

Insert

 the bulb (5) into the thermowell (4) after covering it with 

the paste to improve the transmission of heat.  

Verify

 that the bulb touches the bottom.  

Insert

 the PTFE seal with the relative stainless steel washers into 

the fitting (3).  

Screw

 the seal press (2) onto the fitting  (3) taking care not to 

bend the capillary and relative sheath and tighten until the PTFE 
seal is tight on the capillary tube.  

Run

  the  capillary  protected  by  the  armor  in  the  established 

direction, avoiding tight bends, and block using the stainless steel 
bands. If a large amount of capillary remains this should be rolled 
up and fixed tightly. The coil must not have a diameter of less 
than 200 mm. 

9.4  TEMPERATURE  SWITCHES  WITH  HELICAL 
BULB 

The temperature switch has to be installed protecting the sensing 
element (bulb) against heat radiation and accidental bumps . 

Fig. 6 -   

Temperature switch for ambient temperature measurement

 

 

 

9.5  ELECTRICAL CONNECTIONS 

It  is  recommended  to  carry  out  the  electrical  connections 
according to the applicable standards. In case of flameproof or 
intrinsic  safety  instruments  see  also  the  standards  IEC/EN-
60079-14. The following mounting arrangements are possible. 

9.5.1 MOUNTING WITH CORD CABLE 

Run the cable so that it cannot be easily damaged (e.g. due to 
too narrow bends, heat sources) and strain it.  
Mount, if provided, the external ground device on the electrical 
connection of the instrument.  
The  external  ground  screw  is  obligatory  for  flameproof 
constructions. 
This  device  is  to  be  threaded  on  while  holding  the  electrical 
connection steady with a 27 mm wrench on hex, until it reaches 
the bottom of the thread (Fig. 7). 

Fig. 7 - Mounting of ground device  

 

DO NOT TURN 

 

9.5.2 MOUNTING  WITH  CABLE  PROTECTED  BY 
FLEXIBLE TUBING 

Mount, if provided, the external ground device on the electrical 
connection of the instrument. This device is to be threaded on, 
while  holding  the  electrical  connection  steady  with  a  27  mm 
wrench on hex, until it reaches the bottom of the thread (fig. 8). 
The  external  ground  screw  is  obligatory  for  flameproof 
constructions. 
If the flexible tubing is fitted with a male thread, apply a nipple to 
the electrical connection. 

CAUTION:

  the  adaptor  is  to  be  threaded  on  while  holding  the 

electrical connection steady with a 27 mm wrench on hex (fig.8)  

 

Fig. 8 - Mounting of adaptors 

 

DO NOT TURN 

 

9.5.3 MOUNTING  WITH  CABLE  PROTECTED  BY 
METAL TUBING 

CAUTION:

 mounting with cable protected by metal tubing is to 

be carried out avoiding any torque on the electrical connection 
while assembling the various fittings. 
In  order  to  make  easier  the  instrument  removal  for  check  and 
calibration  it  is  advisable  to  apply  on  the  instrument  a  junction 
box provided with terminals. 

CAUTION:

 accessories used for installation have to be certified 

according to standards EN or IEC 

 

Summary of Contents for TXS Series

Page 1: ...eights are not binding unless released on certified drawings CAUTION Before installing using or carrying out maintenance on the instrument it is necessary to read and understand the indications given...

Page 2: ...vates one or two simultaneous release electric microswitches The microswitches are of the rapid release type with automatic rearm When the temperature moves away from the set values returning towards...

Page 3: ...strument closes the circuit when the set point is reached with temperature on fall closure on fall The temperature switch must be kept in the normal installation position i e with the temperature conn...

Page 4: ...ion and accidental bumps Fig 6 Temperature switch for ambient temperature measurement 9 5 ELECTRICAL CONNECTIONS It is recommended to carry out the electrical connections according to the applicable s...

Page 5: ...external condition of the enclosure In case of flameproof or intrinsic safety instruments inspections of the electrical installation are to be carried out also according to customer procedures and at...

Page 6: ...ened electrical joints Interrupted or short circuited electrical line Microswitch contacts damaged Filling fluid leakage Check all electrical joints Check the conditions of the electrical line Replace...

Page 7: ...allation Annex 1 Model code 1 MODEL CODE TX For further information see datasheet 1 1 Ignition protection mode S Weatherproof A Ex d 1 2 Enclosure code material 2 Aluminium 4 316 s s B Direct Mounting...

Reviews: