background image

 

INSTRUCTION MANUAL 

NI-492WE 

Rev.4  05/18 

 

3 of 7 

The position of the adjusting bush is given in Fig. 2. 

Fig. 2 - Adjusting bush and plumbing 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 Adjustment slot cover 

 Sealing lead  

 

Stainless steel wire 

 Adjusting bush 

The effect of the sense of rotation of the adjusting bush is shown 
on the slot cover. Rotate the bush by inserting a 

3mm

 dia. rod or 

drill into the holes on the bush itself (It recommends using a drill 
bit on the side of the shank). 

7 -  SET POINT CALIBRATION 

In  order  to  proceed  with  the  calibration  and  the  periodical 
functional  verification  of  the  instrument  a  suitable 

calibration 

circuit

 (Fig. 3) and of an adequate source of heat are required.  

The  test  instrument  should  have  a  measurement  range 
approximately equal to or slightly wider than the pressure switch 
range and should have an accuracy consistent with the accuracy 
required to calibrate the set point. 

Fig. 3 – Calibration circuit  

 

TS - 

Temperature switch 

TT - 

Test thermometer 

TB - 

Thermostatic bath 

7.1  PRELIMINARY OPERATIONS  

Slide up the adjustment cover (Fig. 2 item 1). 

7.2  CALIBRATION CIRCUIT AND OPERATIONS 

Prepare the control circuit as indicated in Fig.3

 

If the instrument is equipped with two contacts, remember that 
they  are  released  simultaneously  but  within  the  specification 
tolerance. 
The warning lamps should be connected to the terminals NA or 
NC according to the required contact action. 

Fig. 4 – Color of wires and function 

GDN-Internal grounding connection: color jellow/green

 

C-Common: 

color brown 

NO-Normally open: 

color blue

 

NC-Normally closed: 

color black

 

M

icro 

C-Common: 

color gray 

NO-Normally open: 

color red

 

NC-Normally closed: 

color white

 

M

icro 

 

Contacts status at atmospheric pressure. 

 

C

  -  Common 

NA

 - Normally open 

NC

 - Normally closed   

Connection between C and NA

 

• 

If the circuit is open at the workin temperature, the instrument 

closes

 the circuit when the set point is reached with temperature 

on rise

 (

closure on rise

). 

 If the circuit is closed at the working temperature, the instrument 

opens

 the circuit when the set point is reached with temperature 

on

 

fall

  (

opening on fall

). 

Connection between C and NC  

 If the circuit is closed at the working temperature, the instrument 

opens

 the circuit when the set point is reached with temperature 

on

 

rise

 (

opening on rise

).  

• If the circuit is open at the working temperature, the instrument 

closes

 the circuit when the set point is reached with temperature 

on

 

fall

 (

closure on fall

). 

The temperature switch must be kept in the normal installation 
position, i.e. with the temperature connection downwards. 
Modify  the  temperature  in  the  circuit  up  to  the  desired 
microswitch set point value. 
Using  a 

3mm

  dia.  rod  drill  into  the  holes  on  the  bush  (It 

recommends using a drill bit on the side of the shank). 
Adjust its position until the relative lamp turns on (or turns off); 
then turn it in the opposite direction until the lamp turns off (or 
on). Slowly turn the bush again until the lamp turns on (or off) 

7.3  SET POINT VERIFICATION 

Increase the temperature up to the normal working temperature 
(Tw), wait the stabilization of the temperature. Modify, slowly the 
temperature  up  to  the  set  value  (Ti)  and  record  the  set  point 
value. Write the set point value on the adhesive label. 

Note

:  the  verification  of  the  repeatability  have  to  be  tested 

checking the set point (Ti) three times approaching the set point 
from the same starting point (Tw). The temperature cycles have 
to be performed slowly. 

7.4  FINAL OPERATIONS 

Disconnect the instrument from the calibration circuit. 
Close the adjustment slot by sliding down the slot cover (Fig. 2, 
1), then seal with lead the instrument. 
Mount on pressure connection the protecting cap supplied with 
the instrument. It should be definitively removed only 

during

 the 

connection steps.  

8 -  INSTRUMENT PLUMBING 

The  plumbing  (see  Fig.  2),  aimed  as  a  guarantee  against 
possible tampering of the calibration, can be carried out using a 
flexible  steel  wire  (2),  1  mm

2

  in  section,  wound  up  around  the 

case in the groove purposely provided. 

9 -  MOUNTING AND CONNECTIONS 

9.1  MOUNTING 

Mount  instruments  with  capillary  either  on  pipe  or  surface  by 
means of the proper bracket (see Fig. 9 and 10) or direct on plant.  
Select  a  location  where    possible  shocks  and  temperature 
variations remain within tolerable limits. 
For the instrument with capillary the difference in height between 
bulb  and  case  have  not  exceeding  two  meters  (fig  11  and  12 
distance h). 

9.2  INSTALLATION OF TEMPERATURE SWITCH 
WITH DIFFERENCE IN HEIGHT BETWEEN BULB 
AND INSTRUMENT ENCLOSURE GREATER THAN 2 
METERS 
9.2.1  TEMPERATURE SWITCHES CLASS SAMA II A 

Difference  in  height  between  bulb  and  instrument  enclosure 
causes  a  systematic  error  of  set  point  calibrated  value  (“bulb 
elevation error”). This error can be corrected during calibration 
using the table attached to our technical instruction 

IS-TC.401E

available on request. 

Summary of Contents for TXS Series

Page 1: ...eights are not binding unless released on certified drawings CAUTION Before installing using or carrying out maintenance on the instrument it is necessary to read and understand the indications given...

Page 2: ...vates one or two simultaneous release electric microswitches The microswitches are of the rapid release type with automatic rearm When the temperature moves away from the set values returning towards...

Page 3: ...strument closes the circuit when the set point is reached with temperature on fall closure on fall The temperature switch must be kept in the normal installation position i e with the temperature conn...

Page 4: ...ion and accidental bumps Fig 6 Temperature switch for ambient temperature measurement 9 5 ELECTRICAL CONNECTIONS It is recommended to carry out the electrical connections according to the applicable s...

Page 5: ...external condition of the enclosure In case of flameproof or intrinsic safety instruments inspections of the electrical installation are to be carried out also according to customer procedures and at...

Page 6: ...ened electrical joints Interrupted or short circuited electrical line Microswitch contacts damaged Filling fluid leakage Check all electrical joints Check the conditions of the electrical line Replace...

Page 7: ...allation Annex 1 Model code 1 MODEL CODE TX For further information see datasheet 1 1 Ignition protection mode S Weatherproof A Ex d 1 2 Enclosure code material 2 Aluminium 4 316 s s B Direct Mounting...

Reviews: