16
Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
Display
Bar graph display
Unit
Full scale
Max. value
Min. value
Main display
Special function
2nd display
Tare
Battery-charge-level
indicator LO-BAT
Min./max. memory
The min./max. memory is reset to the current measured value with each measuring cycle.
■
The display of the min. value is achieved by pressing the Min key.
■
The display of the max. value is achieved by pressing the Max key.
■
The resetting of the current measured value is achieved by holding down the respec-
tive key (Min or Max) and additionally briefly pressing the Menu key.
Bar graph with drag-pointer function
The additional integrated bar graph with drag-pointer function within the display
indicates the trend of the current operating pressure directly, independently from
the digital indicator. Using the drag-pointer function, the max. memory value is also
displayed in the bar graph display in the form of a bar segment, as well as in the digital
display. On resetting the min./max. value, this bar graph is then updated to the current
measured value.
Voltage supply
The digital pressure gauge is operated with 2 AA-Mignon batteries. The status of the
batteries is shown in the display . With a low battery charge level, a blinking battery
symbol is shown on the indicator display - the batteries should then be changed as soon
as possible.
■
Open the battery cover of the digital pressure gauge on the rear of the instrument.
■
Replace the batteries with 2 new mignon cells.
■
Close the battery cover again.
5. Commissioning, operation
For further information and advice see the appropriate operating
instructions DG-10.