11
Operating instructions, models F6137, F6148, F6160, F6171
ADPR1X914115.01 10/2020 EN/DE
EN
4. Transport, packaging and storage
4. Transport, packaging and storage
4.1 Transport
Check the hydraulic ring force transducer for any damage that may have been caused
by transport. Obvious damage must be reported immediately.
CAUTION!
Damage through improper transport
With improper transport, a high level of damage to property can occur.
▶
When unloading packed goods upon delivery as well as during
internal transport, proceed carefully and observe the symbols on the
packaging.
▶
With internal transport, observe the instructions in chapter 4.2
“Packaging and storage”.
As precision measuring instruments, hydraulic ring force transducers require careful
handling during transport and mounting. Load impacts during transport (e.g. hitting
a hard surface) can lead to irreversible damage, resulting in measuring errors in the
subsequent measuring operation.
4.2 Packaging and storage
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change
in installation site, sending for repair).
The hydraulic ring force transducer has been manufactured from coated steel.
■
The design with pressure gauge fulfils IP65 ingress protection.
■
The design with digital pressure gauge fulfils IP65 ingress protection.
■
The design with pressure sensor fulfils IP67 ingress protection. The design of the
accessory cables fulfils IP67 ingress protection. The ingress protection IP67 is only
guaranteed in the plugged-in state.
During storage the protective cap must always be on the electrical connection to avoid
the ingress of moisture and dirt.
Permissible conditions at the place of storage:
■
Storage temperature: -30 ... +60 °C
■
Humidity: 35 ... 85 % relative humidity (non-condensing)
Avoid exposure to the following factors:
■
Mechanical vibration, mechanical shock (putting it down hard)
■
Dust, dirt, and other objects may not be deposited in such a way that they form a
force shunt with the measuring piston, since this will falsify the measurement.