26
WIKA operating instructions, models CTH63I0 and CTH65I0
EN
14120692.02 01/2019 EN/DE
9. Enter the temperature of your reference now.
10. Change the digits using the arrow key
[▲]
.
11. Use the arrow key
[▼]
to switch to the next digit.
12. Confirm by pressing
[ENTER/MENU]
after completing the input of the temperature.
13. Press
[ESC]
to return to the measuring mode.
The physical calibration
oP2
cannot be cancelled by means of the
[ESC]
key.
If required, the calibration can be cancelled by switching off the measuring
instrument.
Example of the measured value display after physical calibration against a
reference standard:
The small
2
in the left corner in connection with the display segment
CAL
in the centre of
the display indicates that the physical probe calibration
oP2
has been activated.
6.4.4 Activating/deactivating measuring channels [Chnl]
(only for 2-channel instruments)
Chnl = Channel = select
1. Press the
[ENTER/MENU]
key and select
Chnl
using the arrow keys
[▲]
or
[▼]
.
2. Press
[ENTER/MENU]
key once again.
⇒
A small 1 appears on the left side of the display, it indicates the channel.
3. Use the arrow keys
[▲]
or
[▼]
to select the channel to be activated or deactivated.
4. Confirm selection by pressing
[ENTER/MENU]
.
5. Use the arrow keys
[▲]
or
[▼]
now to activate
on
or deactivate
off
the set channel.
6. Press
[ENTER/MENU]
to confirm the desired setting.
7. Use
[ESC]
to return to the measuring mode.
6. Commissioning, operation