27
WIKA operating instructions, models CTH63I0 and CTH65I0
EN
14120692.02 01/2019 EN/DE
Alternative: keep the
[HOLD/MAX/MIN/AVE]
key pressed for 2 seconds; this deactivates
or activates channel 2.
At least one channel is always active!
6.4.5 Storage management [Lo6] (not possible for CTH63I0)
1. Press the
[ENTER/MENU]
key and select
Lo6
using the arrow keys
[▲]
or
[▼]
.
⇒
Only when the data logger version has been ordered.
2. Press
[ENTER/MENU]
key once again.
⇒
OFF
appears in the lower display line.
3. Now use the
[▲]
or
[▼]
arrow keys to start the logger mode with
ON
.
4. Confirm selection by pressing
[ENTER/MENU]
.
5. Use the arrow keys
[▲]
or
[▼]
to choose between the automatic
Auto
or manual
SPot
saving.
6. Confirm by pressing
[ENTER/MENU]
.
7. Decide whether to “Add” or “Create” a new log file by selecting
Add
and
nLo6
.
⇒
When selecting the automatic saving at the end you are asked to select the
measuring interval:
1S
= 1 second
1M
= 1 minute
5S
= 5 seconds
2M
= 2 minutes
10S = 10 seconds
5M
= 5 minutes
20S = 20 seconds
10M = 10 minutes
30S = 30 seconds
20M = 20 minutes
8. Confirm the input by pressing
[ENTER/MENU]
.
9. Use
[ESC]
to return to the measuring mode.
MENU
Unit
Lin2
CAL
Chnl
Lo6
MENU
Unit
Lin2
CAL
Chnl
Lo6
6. Commissioning, operation