23
WIKA operating instructions, models CTH63I0 and CTH65I0
EN
14120692.02 01/2019 EN/DE
3. Use the
[▲]
or
[▼]
arrow keys to select the channel (
1
or
2
) to be calibrated.
⇒
Channel selection only possible with 2-channel instruments.
4. Confirm selection by pressing
[ENTER/MENU]
.
5. Now use the arrow keys
[▲]
or
[▼]
to select the desired calibration function.
Standard linearisation in accordance with DIN IEC 60751 “CoFF”
1. Use the arrow keys
[▲]
or
[▼]
to select
CoFF
.
2. Press
[ENTER/MENU]
to confirm the desired setting.
3. Use
[ESC]
to return to the measuring mode.
Calibration by code oP1
1. Use the
[▲]
or
[▼]
arrow keys to select
oP1
.
2. Press
[ENTER/MENU]
to confirm the desired setting.
⇒
A small 1 appears in the lower part of the display. 4 characters follow it (hex code /
0 ... F).
3. The 4 characters can be changed upwards by means of the
[▲]
arrow key.
4. Press the
[▼]
arrow key to go to the next position.
5. Confirm by pressing
[ENTER/MENU]
if all 4 characters are entered as required.
⇒
A small 2 appears and the following 4 characters can be changed as well.
6. Use
[ESC]
to return to the measuring mode.
MENU
Unit
Lin2
CAL
Chnl
Lo6
CAL
CoFF
oP1
oP2
6. Commissioning, operation