29
WIKA operating instructions, model CTD4000
14366831.01 05/2020 EN/DE
EN
12. Specifications
Specifications
Model CTD4000-375
Model CTD4000-650
Display
Temperature range
t
amb
+ 15 °C ... 375 °C [t
amb
+ 15 °F ... 707 °F]
t
amb
+ 15 °C ... 650 °C [t
amb
+ 15 °F ... 1,202 °F]
Accuracy
1)
0.35 K
0.5 K
Stability
2)
±0.1 K
±0.3 K
Resolution
0.1 °C
Temperature control
Heating time
approx. 20 min from 30 to 375 °C
[from 86 °F to 707 °F]
approx. 35 min from 50 to 650 °C
[from 122 °F to 1,202 °F]
Cooling time
approx. 40 min from 375 to 100 °C
[from 707 °F to 212 °F]
approx. 60 min from 650 to 100 °C
[from 1,202 °F to 212 °F]
Stabilisation time
3)
dependent on temperature and temperature probe
Insert
Immersion depth
150 mm [5.91 in]
Insert dimensions
Ø 26 x 150 mm [Ø 1.02 x 5.91 in]
Insert material
Aluminium
Brass, nickel-plated
Voltage supply
Operating voltage
AC 115/230 V ±10 %, 50/60 Hz
Automatically switchable
Power consumption
600 W
Fuse
6.3 A slow blow fuse (at AC 115 V)
3.15 A slow blow fuse (at AC 230 V)
Power cord
AC 230 V; for Europe
Communication
Interface
RS-232
Case
Dimensions (W x D x H)
130 x 260 x 280 mm [5.12 x 10.24 x 11.02 in]
Weight
5.4 kg [11.9 lbs]
6 kg [13.2 lbs]
1) Is defined as the measuring deviation between the measured value and the reference value.
2) Maximum temperature difference at a stable temperature over 30 minutes.
3) Time before reaching a stable value.
The measurement uncertainty is defined as the total measurement uncertainty (k = 2), which contains the following shares:
Accuracy, measurement uncertainty of reference, stability and homogeneity.
Certificates
Certificate
Calibration
■
Without
■
3.1 calibration certificate per DIN EN 10204
■
DKD/DAkkS calibration certificate
Recommended recalibration interval
1 year (dependent on conditions of use)
Approvals and certificates, see website
For further specifications see WIKA data sheet CT 41.10 and the order documentation.