9
WIKA operating instructions, model CTD4000
14366831.01 05/2020 EN/DE
EN
4. Design and function
4.4 Voltage supply
13
Information about the calibrator
14
Main switch
15
Fuse
16
Power connection
17
RS-232 interface
For further information, see chapter 8 “Serial
communication”.
Voltage supply for CTD4000-140
The CTD4000-140 calibrator operates with an operating
voltage of AC 100 ... 240 V, 50/60 Hz.
Voltage supply for CTD4000-375 and CTD4000-650
The calibrator runs on a voltage of AC 230 V or AC 115 V,
50/60 Hz.
The calibrator automatically sets the supply voltage to a
voltage of AC 115 V or AC 230 V.
The instruments can be supplied with voltage of AC 115 V or
AC 230 V (50/60 Hz).
The fuse must be changed, if the supply voltage changes
from AC 230 V to AC 115 V.
Voltage
Fuse
AC 230 V
3.15 A (factory delivery)
AC 115 V
6.3 A (included in standard scope of delivery, in
a bag)
For both calibrators, in total 4 fuses are included in the
standard scope of delivery. One is already built-in; the
remaining are placed in labelled plastic bags.
13
14
15
16
17
4.5 Fuse
The calibrators are fitted with the following fuses, in
accordance with the table:
Operating
voltage
CTD4000-140
CTD4000-375
CTD4000-650
AC 100 ... 240 V
2.5 A slow blow
fuse
-
AC 100/115 V
-
6.3 A slow blow
fuse
AC 230 V
-
3.15 A slow blow
fuse
Only use type F 5 x 20 mm fuses. All the
electrical components are found below the main
switch.
Changing of the fuse
1. Open the fuse container, using a screwdriver for example,
and pull it out.
⇒
2. Remove the fuse from the clamp.
3. Place new fuse into the clamp.