35
WIKA operating instruction, model CPP120-X
EN
08/2020 EN/DE
10. Specifications
10. Specifications
Specifications
Model CPP120-X
Pressure range
0 ... 120 bar
[0 ... 1,740 psi]
Permissible pressure supply
Pressure supply
100 % of test item pressure range, max. 120 bar [1,740 psi]
Pressure transmission
medium
Clean, dry and non-corrosive gases (e.g. Air or Nitrogen)
Pressure connections
Test connections
2 x G 1 female thread incl. 2 x test item connections with G ½
female thread, loose union connection
Distance between test
connections
288 mm [11.34 in]
External pressure supply
connection
G ¼ B female thread
Fine adjustment
Volume adjuster
Material
Wetted parts
Austenitic stainless steel, high tensile brass, copper, nitrile
rubber
Case
ABS material
Permissible ambient conditions
Operating temperature
0 ... 40 °C [32 ... 104 °F]
Storage temperature
-10 ... +50 °C [14 … 122 °F]
Instrument base
Dimensions (W x D x H)
510 x 490 x 210 mm [20.08 x 19.29 x 8.27 in], for details, see
technical drawings
Weight
22 kg [48.51 lbs]
For further specifications see WIKA data sheet CT 91.03 and the order documentation.