24
WIKA operating instruction, model CPP120-X
EN
08/2020 EN/DE
6. Commissioning, operation / 7. Faults
Instructions
1. Fill syringe with solvent (suitable cold degreasing liquid).
2. With gauge connection pointing upwards put needle into connection and insert by feel
the point into the hole leading to the tube.
3. Inject the solvent. Ideally the tube should be half full.
4. Shake gauge in various attitudes to agitate solvent.
5. Suck solvent back into syringe, holding gauge at an angle.
6. Check that solvent removed is clean. To be sure that all oil has been removed, repeat
cleaning process until solvent removed from gauge is as clean as that put in.
Cleaning of gauges
Filling with solvent
Removing the
solvent
7. Faults
Personnel:
Skilled personnel
Protective equipment:
safety goggles
Tools:
open-end wrench
Only use original parts (see chapter 11 "Accessories").
For contact details, please see chapter 1 "General information" or the back
page of the operating instructions.