6. Commissioning, operation
6. Commissioning, operation
Personnel:
Skilled personnel
Protective equipment:
safety goggles
Tools:
open-end wrench
Only use original parts (see chapter 11 "Accessories").
6.1 Unpacking the pneumatic comparison test pump
As soon as possible after delivery open the packaging of the model CPP120-X pneumatic
comparison test pump and check that all the items detailed in the packing list (see
chapter 2.3 "Scope of delivery") are included.
Unpacking the items, check them for any damage that may have been caused by
transport. If any items are missing, please contact DH-Budenberg/WIKA immediately.
6.2 Environmental requirement
When not sitting the CPP120-X in a temperature controlled laboratory look for an area
that satisfies the following criteria as much as possible:
■
A constant temperature area free from draughts and supplies of heat or cold
■
An area free from noise and vibration, constantly used pathways
■
A clean dry area free from corrosive liquids or vapours
A strong, stable, level table or workbench with the capability of supporting the system with
sufficient space to operate is required.
6.3 Assembly of comparison test pump
Fastening the CPP120-X to the bench
The CPP120-X should be mounted on a firm, level table or bench about 0.9 m [35.4 in]
high. The centre line of the front adjustable feet of the CPP120-X should be about 40 mm
from the front edge of the bench to allow adequate clearance for the volume adjuster
5
.
1. Fit the pneumatic comparison test pump on the bench with the adjustable feet on the
level plates and the volume adjuster
5
projecting over the front of the bench.
2. Using the spirit level provided and level the unit in both the front/rear axis and the side
to side axis by adjusting the four knurled feet.