7. Faults
CAUTION!
Physical injuries and damage to property and the environment
If faults cannot be eliminated by means of the measures listed, the
comparison test pump must be taken out of operation immediately.
▶
Ensure that pressure is no longer present and protect against accidental
commissioning.
▶
Contact the manufacturer.
▶
Wear the requisite protective equipment (see chapter 3.5 "Personal
▶
If a return is needed, please follow the instructions given in chapter
Faults
Causes
Measures
Test pump does not
build up pressure
Pressure supply not connected
Check whether pressure supply is
connected and gas cylinder is full.
Connection to inlet manifold
leaking
Check for leaks with soapy water.
Blocked inlet valve
Check the seals.
If necessary return to
manufacturer.
Test connection leaking
Check the seals.
Replace if necessary.
Inlet valve handle is loose
Check whether the screw has
loosened, tighten the screw if
necessary.
Incorrect operation
Ensure that correct operating
procedure is being followed
(see chapter 6 "Commissioning,
If unable to locate a cause
Return to manufacturer.
Test pump build up
pressure, but pressure
drops down again
The needle valve of the outlet
valve does not seal.
⇒
Recognizable by hissing
sound
Tighten outlet valve again.
Check the seals.
Replace if necessary.
Missing or damaged seals.
⇒
Recognizable by hissing
sound
Check the sealings for correctly
seating and wear.
Replace if necessary.
If unable to locate a cause
Return to the manufacturer.