Mini-Seis III Pro™ Operating Manual
Page 26
After 4 hours have elapsed, the histogram will close and a new histogram will start. The
process will continue until the histogram is made inactive.
T
O
R
UN AN
I
NTERVAL
H
ISTOGRAM
S
TARTING AT A
S
PECIFIC
T
IME
1.
Go to Timer Setup, set the On Timer to Enabled and set the On time HR and MIN to
the time to start.
2.
Set the rate and period as desired.
3.
Set the Interval to the number of hours desired. For illustration we will use 4.
The mode will activate automatically when the set Start Time is reached. After 4 hours has
elapsed, the histogram will close and a new histogram will start. The process will continue
until the histogram is made inactive. The Start Time will not be reused.
T
O
R
UN AN
I
NTERVAL
H
ISTOGRAM
S
TARTING AND
E
NDING AT A
S
PECIFIC
T
IME
1.
Go to Timer Setup, set the On and Off Timer to Enabled and set the On and Off HR
and MIN to the time to start and end.
2.
Set the rate and period as desired.
3.
Set the Interval to the number of hours desired. For illustration we will use 4.
The mode will activate automatically when the set On time is reached and inactivate
automatically when the Off time is reached. If the Off time is more than the Interval, after the
Interval has elapsed, the histogram will close and a new histogram will start. If the Off time
is less than the Interval, the histogram will close at the Off time. If the seismograph is not
manually inactivated, the process will continue each day until the histogram is made
inactive.
Status
To manually activate the histogram select the Activate soft key. The soft key text will change
to Inactivt. To end an active histogram select the Inactivt soft key. A histogram cannot be
manually activated if the Timer is set.
Note: Do not turn off power to the instrument while the
histogram is active. Making the histogram inactive allows the
firmware to write the proper summary entries. Failure to do
this may result in a corrupted record.