Error Code

Cause

Corrective Steps

Last run 

memory not 

displayed

Turning the centrifuge ON 

immediately after turning it OFF

Maintain a three second gap 

between turning the centrifuge 

OFF and ON again.

System 

locks up*

Electronics error

Turn the centrifuge OFF and 

then ON. If the error persists, 

contact the technical support 

team.

*  If the system locks up or overheats due to over current, turn the centrifuge OFF and 

then ON again.

ATTENTION:

 If the motor overheats, there will be a fluctuation in speed value to allow 

the centrifuge to cool down for at least 30 minutes. Do not use the unit for 30 minutes.

 

8. Product Recycling

In the case that the product is to be disposed of, the relevant legal regulations are  

to be observed.

Summary of Contents for 4500RPM Max

Page 1: ...1 PRODUCT MANUAL Multipurpose Lab Centrifuge 4500RPM Max Code 665 113...

Page 2: ...ion the unit can halt and resume operations without disturbing fragile samples To protect the samples further the centrifuge s intelligently designed airflow system ensures that the temperature of the...

Page 3: ...mum Rotor Capacity 4 x 100ml Maximum Speed RPM RCF g 4500 RPM g varies per rotor Speed Setting Range 500 to 4500 RPM Step of 100 RPM Speed Accuracy 100 RPM Time Setting Range 1 to 99 minutes and infin...

Page 4: ...or the rotor chamber the centrifuge must be cleaned carefully before resuming operations The centrifuge may be used for the specified applications only It must not be operated in a hazardous or flamm...

Page 5: ...tamination caused by aggressive agents the rotor must be cleaned immediately with a natural cleaning liquid Pay particular attention to the bores of the tubes If any damage is seen contact the service...

Page 6: ...ed as a lid open function of the centrifuge motor is not running 3 RPM RCF Single press to display the RPM and for the RPM LED to blink Long press the button to toggle the display between RPM and RCF...

Page 7: ...N the power switch located on the right of the unit ATTENTION Maintain a gap of three seconds between turning the power switch OFF and then turning it ON again Do not turn ON and OFF again immediately...

Page 8: ...ters Speed Time ACC DCC as described above To select the programs press the increment and decrement buttons All the parameters in the program will save automatically after the display stops blinking T...

Page 9: ...15ml 4500 3485 L30 140 Rotor 25 32 x 10ml 4500 3164 L30 124 Rotor 26 2 x Microplates 4500 2377 Fixed Angle Rotors Rotor Code Rotor Max Volume Max RPM Max RCF L30 136 Rotor 7 4 x 100ml 4500 2559 L30 1...

Page 10: ...osing and opening mechanism Contact the technical support team on 1800 358 101 Err 55 Rotor loaded asymmetrically Load the rotor symmetrically and restart the centrifuge Lid cannot be opened Spinning...

Page 11: ...ed States Canada Error Code Cause Corrective Steps Last run memory not displayed Turning the centrifuge ON immediately after turning it OFF Maintain a three second gap between turning the centrifuge O...

Page 12: ...Westlab in every discovery...

Reviews: