background image

• 

7

PRODUCT MANUAL

1800 358 101

[email protected]

www.westlab.com.au

+64 9 553 3677

[email protected]

www.westlab.co.nz

646 568 5391

[email protected]

www.westlab.com

1 877 822 1455

[email protected]

www.westlab.com

Australia

New Zealand

United States

Canada

Number

Name

Description

8

PEdit

Single press for the Speed/Time/ACC/DCC to blink 

one after another five times on respective display. The 

user can change the blinking parameters by using the 

increment and decrement buttons. In PEdit mode, the 

user can also press the Speed/Time/ACC/DCC and 

change its value.

9

PROG

Single press to select the program from one to 99. The 

different parameters can be set in each program for 

quick selection.

10

Quick Spin

Press and hold for immediate centrifugation up to the 

last selected RPM.

Turning ON the centrifuge

ATTENTION: 

Check the rotor fitment before use. The centrifuge will not operate with  

an open lid.

1.  Ensure that the power switch is OFF when connecting the power cord.
2.  Connect the power cord to the centrifuge.
3.  Turn ON the main power supply and then turn ON the power switch located  

on the right of the unit.

ATTENTION: 

Maintain a gap of three seconds between turning the power switch OFF 

and then turning it ON again. Do not turn ON and OFF again immediately. 

Speed setting

Press the RPM/RCF button to set the speed from 500 to 4500 RPM. Single press to 

set the speed value and long press the button to toggle the display between RPM 

and RCF modes. The parameter can be set when the speed display blinks by using 

the increment and decrement buttons. The values are automatically saved after the 

display stops blinking.

Timer setting

Press the timer button to set the time value. The time value can be set by using the 

increment and decrement buttons. The value will be automatically saved after the 

display stops blinking. The timer can only be set in minutes while the seconds are for 

the countdown timer.

Acceleration and deceleration 

The ACC/DCC button will set the parameters of acceleration and deceleration of the 

rotor ramp. Single press the ACC/DCC button to set the acceleration from one to nine 

by using the increment and decrement buttons. Alternatively, press the ACC/DCC 

once to set the deceleration from one to nine by using the increment and decrement 

buttons. The values of acceleration and deceleration are listed below:

Summary of Contents for 4500RPM Max

Page 1: ...1 PRODUCT MANUAL Multipurpose Lab Centrifuge 4500RPM Max Code 665 113...

Page 2: ...ion the unit can halt and resume operations without disturbing fragile samples To protect the samples further the centrifuge s intelligently designed airflow system ensures that the temperature of the...

Page 3: ...mum Rotor Capacity 4 x 100ml Maximum Speed RPM RCF g 4500 RPM g varies per rotor Speed Setting Range 500 to 4500 RPM Step of 100 RPM Speed Accuracy 100 RPM Time Setting Range 1 to 99 minutes and infin...

Page 4: ...or the rotor chamber the centrifuge must be cleaned carefully before resuming operations The centrifuge may be used for the specified applications only It must not be operated in a hazardous or flamm...

Page 5: ...tamination caused by aggressive agents the rotor must be cleaned immediately with a natural cleaning liquid Pay particular attention to the bores of the tubes If any damage is seen contact the service...

Page 6: ...ed as a lid open function of the centrifuge motor is not running 3 RPM RCF Single press to display the RPM and for the RPM LED to blink Long press the button to toggle the display between RPM and RCF...

Page 7: ...N the power switch located on the right of the unit ATTENTION Maintain a gap of three seconds between turning the power switch OFF and then turning it ON again Do not turn ON and OFF again immediately...

Page 8: ...ters Speed Time ACC DCC as described above To select the programs press the increment and decrement buttons All the parameters in the program will save automatically after the display stops blinking T...

Page 9: ...15ml 4500 3485 L30 140 Rotor 25 32 x 10ml 4500 3164 L30 124 Rotor 26 2 x Microplates 4500 2377 Fixed Angle Rotors Rotor Code Rotor Max Volume Max RPM Max RCF L30 136 Rotor 7 4 x 100ml 4500 2559 L30 1...

Page 10: ...osing and opening mechanism Contact the technical support team on 1800 358 101 Err 55 Rotor loaded asymmetrically Load the rotor symmetrically and restart the centrifuge Lid cannot be opened Spinning...

Page 11: ...ed States Canada Error Code Cause Corrective Steps Last run memory not displayed Turning the centrifuge ON immediately after turning it OFF Maintain a three second gap between turning the centrifuge O...

Page 12: ...Westlab in every discovery...

Reviews: