Page 6 of 14 

Technical data 

Dimensions (W x L x H) 

: 240 (9.44) X 270 (10.63) X 101 (3,97) mm (inches) 

Mains/line voltage 

: 230 V (120 V) AC 

Power input  

: 475 W 

Air volume 

: 5 – 50 l/min. 

Temperature range 

: 50°C – 550°C 

Accuracy 

:  + - 30°C (+ - 54°F) 

Vacuum 

: 0.6 bar 

Compressed air connection 

: 400 -  600 kPa 

Main fusing 

: 230 V / T3,15 A  (120 V / T6,3 A)  

Protection class  

: 1 (control unit and handpiece hard grounded)  

 
 

2. Placing in operation 

Place the hand-piece with the hot air nozzle installed in the safety stand AKT 30. (Without the safety 
stand or optional WBH 30000S Circuit Board Holder the unit cannot be placed in operation.)  
 
For operation of the WHA 3000V, clean, dry compressed air or, alternatively, nitrogen N

2

 is required. 

The compressed air connections (18, 19) are on the rear side of the unit. For connection of the 
compressed air supply, a compressed air hose with an outside diameter of 6 mm, which is suitable 
for operating pressure, is required.  
 
The nitrogen connection (19) serves to generate hot air. 
The compressed air connection (18) serves to generate vacuum by means of a compressed air 
converter.  
 
If a separate nitrogen supply is not used, the compressed air connections (18) and (19) can be 
linked to the Y-connector and supplied with normal compressed air. 
 
Plug cable from safety stand or optional WBH 3000S Circuit Board Holder (proximity switch), into 
socket (20). Check whether the mains voltage matches the information on the rating plate. If the 
mains voltage is correct, connect the controller to the mains. When the device is switched on, the 
device name 

"WHA 3000" 

and the software 

"Version"

 are displayed briefly. The electronics then 

switch automatically to the default setting (Manual operating mode). 
 
 

2.1 Manual operating mode 

 
 
Information on 
the display(1) 
 
 
 
 
 

 Nozzle 

Air 

Nozzle 

 

Temp. 

Flow Rate 

Pre Heating 

 

in °C 

in l/min 

ON/OFF 

 
In the Manual operating mode, the parameters for nozzle temperature, air flow rate and Switching 
on or off the pre- heating can be set. 
 
 
 
 

 

TEMP AIR 

PREHEAT

 

 

 

 

 

350 °C 

25 l 

OFF 

 

 

 

 

Factory setting 

Summary of Contents for WHA 3000VS

Page 1: ...Weller Tools GmbH Carl Benz Str 2 74354 Besigheim Germany Tel 49 0 7143 580 0 Fax 49 0 7143 580 108 WHA 3000V Hot Air Station Operating Instructions Version 2 3...

Page 2: ...m function 10 2 6 Saving and loading temperature time profiles 10 3 External sensor 11 3 1 External sensor with measuring function MEASURE MODE 11 3 2 Teach in Mode procedure 11 3 3 External sensor wi...

Page 3: ...ol key time specification auto mode preheating temperature 6 AIR Control key air volume 7 TEMP Control key hot air temperature 8 LED Indicator Vacuum 9 LED Indicator START STOP 10 START STOP Control k...

Page 4: ...erface 15 PC Interface RS232 16 Connection for External Sensor Type K 17 Hand piece Connecting Hose 18 Connection for Compressed air 19 Connection for Nitrogen N2 400 600k Pa 20 Stand Switch Connectio...

Page 5: ...ities for use of this repair station The hot air temperature at the hand piece can be adjusted within the range of 50 C 550 C 122 F 999 F Through an external compressed air supply or nitrogen N2 as an...

Page 6: ...ing pressure is required The nitrogen connection 19 serves to generate hot air The compressed air connection 18 serves to generate vacuum by means of a compressed air converter If a separate nitrogen...

Page 7: ...air flow rate 200 C 392 F 5l min Note Hot air is automatically disabled after 999 sec for safety reasons 2 2 Automatic operating mode The Automatic operating mode has a 3 step temperature time profil...

Page 8: ...ce Nozzle temperature during the first step Depressed Twice Temperature of the bottom heater during the first step only with WHP3000 Repeat for Steps 1 3 Symbol for nozzle or bottom heater indicates w...

Page 9: ...from the Locked center position It must remain in position directly over the Hot Plate to allow for the WHA 3000V to operate 2 5 Control of the vacuum function The nozzle vacuum are activated Directl...

Page 10: ...riefly in the display 1 THERMOCOUPLE or THERMOCOUPLE MEASURE MODE CONTROL MODE Both operating modes only become active once the external sensor 16 has been connected 3 1 External Sensor with measuring...

Page 11: ...y be operated using the START STOP 10 and VAC 11 Control Keys 5 Switching Temperature Conversion By pressing the Control Key Temp 7 when switching on the unit 4 you can toggle between the operating mo...

Page 12: ...C LOCKED WHA locked lock function Unlock WHA TOOL STAND AKT30 Stand or WBH3000 Proximity Sensor Connect to socket 20 not connected 9 Accessories 0053119099 External sensor type K 0 5mm 0058754951 Exte...

Page 13: ...2 sides heated Type ND 005 87 368 39 NQT Hot air nozzle 22 0 x 22 0 mm 005 87 368 41 NQT10 Hot air nozzle 14 8 x 14 8 mm 005 87 368 42 NQT25 Hot air nozzle 18 0 x 18 0 mm 005 87 368 43 NQT Hot air noz...

Page 14: ...etting hot air temperature manual mode Setting hot air temperature and bottom heater for step 1 3 auto mode Setting flow rate manual mode setting flow rate for step 1 3 auto mode Time setting for step...

Reviews: