Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
environment–guidance
Voltage Dips/Dropout
IEC 61000-4-11
0 % UT; 0.5 cycle At 0°,
45°, 90°, 135°, 180°, 225°,
270° and 315° % UT; 1
cycle
70 % UT; 25/30 cycles for
50 Hz and 60Hz,
respectively
Single phase: at 0°
0 % UT; 250/300 cycle for
50 Hz and 60 Hz,
respectively
Single phase: at 0°
0 % UT; 0.5 cycle At 0°,
45°, 90°, 135°, 180°, 225°,
270° and 315° % UT; 1
cycle
70 % UT; 25/30 cycles for
50 Hz and 60Hz,
respectively
Single phase: at 0°
0 % UT; 250/300 cycle for
50 Hz and 60 Hz,
respectively
Single phase: at 0°
Mains power quality should be that
of a typical home, commercial, or
hospital environment. If the user of
the RETeval-DR requires continued
operation during power mains
interruptions, it is recommended
that the RETeval-DR be powered
from an uninterruptible power
supply or battery.
Power frequency 50/60Hz
Magnetic field
IEC 61000-4-8
30 A/m, 50 Hz or 60 Hz
30 A/m, 50 Hz or 60 Hz
Power frequency magnetic fields
should be that of a typical home,
commercial, or hospital
environment.
Conducted RF IEC
61000-4-6
3 V, 0.15 MHz – 80 MHz
6 V in ISM radio bands
between 0.15 MHz and 80
MHz
80 % AM at 1 kHz
3 V, 0.15 MHz – 80 MHz
6 V in ISM radio bands
between 0.15 MHz and 80
MHz
80 % AM at 1 kHz
Portable and mobile
communications equipment should
be separated from the RETeval-DR
device by no less than the distances
calculated/listed below:
D= (1.17)
√
P, 150kHz to 80MHz
D= (1.17)
√
P, 80 to 800 MHz
D= (2.33)
√
P, 800 MHz to 2.5 GHz
where P is the max power in watts
and D is the recommended
separation distance in meters. Field
strengths from fixed transmitters, as
determined by an electromagnetic
site survey, should be less than the
compliance levels (V1 and E1).
Interference may occur in the
vicinity of equipment containing a
transmitter.
Radiated RF IEC 61000-4-3 3 V, 0.15 MHz – 80 MHz
6 V in ISM radio bands
between 0.15 MHz and 80
MHz
80 % AM at 1 kHz
3 V, 0.15 MHz – 80 MHz
6 V in ISM radio bands
between 0.15 MHz and 80
MHz
80 % AM at 1 kHz
Portable and mobile
communications equipment should
be separated from the RETeval-DR
device by no less than the distances
calculated/listed below:
D= (1.17)
√
P, 150kHz to 80MHz
D= (1.17)
√
P, 80 to 800 MHz
D= (2.33)
√
P, 800 MHz to 2.5 GHz
where P is the max power in watts
and D is the recommended
separation distance in meters. Field
strengths from fixed transmitters, as
determined by an electromagnetic
site survey, should be less than the
compliance levels (V1 and E1).
Interference may occur in the
Directions for use
Standards and compliance 49
Summary of Contents for RETeval-DR
Page 1: ...Welch Allyn RETeval DR Directions for use...
Page 10: ...6 Equipment serial number Welch Allyn RETeval DR...
Page 14: ...10 About warnings and cautions Welch Allyn RETeval DR...
Page 26: ...22 Setup Welch Allyn RETeval DR...
Page 28: ...24 Operating instructions Welch Allyn RETeval DR...
Page 38: ...34 Test results Welch Allyn RETeval DR...
Page 42: ...38 Clean and disinfect Welch Allyn RETeval DR...
Page 46: ...42 Troubleshooting Welch Allyn RETeval DR...
Page 60: ......