Clean and disinfect
Clean and disinfect
CAUTION
Do not sterilize the device or sensor strips.
CAUTION
Do not submerge the device in liquid or allow liquid to enter the
interior of the device.
CAUTION
Liquid cleaners may damage the LED lights and camera.
CAUTION
Do not use automatic cleansing machines or sterilization.
CAUTION
Consult the cleaning agent and germicidal cleaner agent
manufacturer instructions for their proper use and germicidal efficacy prior
to their use.
CAUTION
Only use the cleaning or germicidal cleaner agent types listed or
damage may occur.
The white interior sphere that the patient looks into (the ganzfeld), should be cleaned
when there is visible dust inside or when the device fails to calibrate at the start of a
test.
Welch Allyn recommends that you clean the eyecup and sensor strip lead between each
patient.
The RETeval-DR device is chemically compatible to wipes containing 70% isopropyl
alcohol and with wipes containing alkyl dimethyl benzyl ammonium chloride. The use of
other wipes may damage the device.
Clean the ganzfeld
1. Remove the eyecup by grasping the rubber nearest the silver bezel and pulling
gently.
2. Clean the ganzfeld with a compressed gas air duster to remove dust.
3. If compressed gas does not work, a damp cloth may be used.
4. To replace the eyecup, orient the eyecup so that the slots on the white plastic rim of
the eyecup are aligned with the bumps on the silver bezel.
5. Push gently until the eyecup clicks onto the silver bezel.
37
Summary of Contents for RETeval-DR
Page 1: ...Welch Allyn RETeval DR Directions for use...
Page 10: ...6 Equipment serial number Welch Allyn RETeval DR...
Page 14: ...10 About warnings and cautions Welch Allyn RETeval DR...
Page 26: ...22 Setup Welch Allyn RETeval DR...
Page 28: ...24 Operating instructions Welch Allyn RETeval DR...
Page 38: ...34 Test results Welch Allyn RETeval DR...
Page 42: ...38 Clean and disinfect Welch Allyn RETeval DR...
Page 46: ...42 Troubleshooting Welch Allyn RETeval DR...
Page 60: ......