Guidance and Manufacturer’s Declaration – Emissions
The RETeval-DR is intended for use in the electromagnetic environment specified below.
The customer or user of the RETeval-DR should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment–
guidance
RF emissions CISPR 11
Group 1
The RETeval-DR device uses RF energy
only for its internal function. Therefore,
its RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class B
Class B
Harmonics IEC 61000-3-2
Class A
Class A
Flicker IEC 61000-3-3
Complies
Complies
The RETeval-DR device is suitable for
use in all establishments, other than
domestic, and those directly connected
to the public low-voltage power supply
network that supplies buildings used
for domestic purposes.
To assure continued effectiveness, only
use cables and accessories supplied by
Welch Allyn which are specifically
designed for use with the RETeval-DR
device.
Guidance and Manufacturer’s Declaration – Immunity
The RETeval-DR device is intended for use in the electromagnetic environment specified
below. The customer or user of the RETeval-DR device should ensure that it is used in
such an environment.
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
environment–guidance
ESD IEC 61000-4-2
±8kV Contact
±15kV Air
±8kV Contact
±15kV Air
Floors should be wood, concrete or
ceramic tile. If floors are synthetic,
the r/h should be at least 30%
EFT IEC 61000-4-4
±2kV Mains
±1kV I/Os
±2kV Mains
±1kV I/Os
Mains power quality should be that
of a typical home, commercial, or
hospital environment
Surge IEC 61000-4-5
±1kV Differential
±2kV Common
±1kV Differential
±2kV Common
Mains power quality should be that
of a typical home, commercial, or
hospital environment
48 Standards and compliance
Welch Allyn
®
RETeval-DR
™
Summary of Contents for RETeval-DR
Page 1: ...Welch Allyn RETeval DR Directions for use...
Page 10: ...6 Equipment serial number Welch Allyn RETeval DR...
Page 14: ...10 About warnings and cautions Welch Allyn RETeval DR...
Page 26: ...22 Setup Welch Allyn RETeval DR...
Page 28: ...24 Operating instructions Welch Allyn RETeval DR...
Page 38: ...34 Test results Welch Allyn RETeval DR...
Page 42: ...38 Clean and disinfect Welch Allyn RETeval DR...
Page 46: ...42 Troubleshooting Welch Allyn RETeval DR...
Page 60: ......