4.7
Slave Error - $02 / XXXXXX1X
4.7.1
Description:
This error occurs only when the Slave has an error of its own that has been written
to the Historical Slave Error Register.
4.7.2
Suggestions:
Verify that the Slave is working properly - Investigate the error indicated in the
Historical Slave Error Register.
4.8
RAM Error - $01 / XXXXXXX1
4.8.1
Description:
This error results from the RAM memory inside the Master microcontroller testing
bad. This memory is used to temporarily store program information while the
Master is operating.
4.8.2
Suggestions:
Replace the Master microcontroller or MCU PCB, clear the error registers, power
down the unit, and re-test the unit.