7
Handle Command List
P H
Ping Handle
Establish communications
C ^A
Read I2E2PROM
Will read one byte from I2E2PROM
C ^B
Write I2E2PROM
Will write one byte to I2E2PROM
C ^C
Pressure Up
Drives pressure up to +200 daPa
C ^D
Pressure Down
Sweeps from +200 daPa to -400 daPa
C ^E
Speaker On
Begins to generate a 226 Hz signal
C ^F
Speaker Off
Turns the speaker off
C ^G
Self-test
Does self test & sends result byte
C ^H
Test I2E2PROM
Tests I2E2PROM, zeroing each location
C ^I
Read Pressure
Returns current pressure reading
C ^J
Read Microphone
Returns current value at mic A/D
C ^K
Start Tymp
Runs a Tymp
C ^L
Set LCD Pattern
Displays a pattern to the LCD (1-7)
C ^M
Close Valve
Closes the valve
C ^N
Auto Zero
Opens the valve and homes the pump
C ^O
Block Rd PROM
Sends the entire contents at once
C ^P
Block Wr PROM
Writes entire contents at once
C ^Q
Calib Mode
Disables time-outs
C ^R
Enable TimeOuts
Enables time-outs
C ^S
Read Button
Read the state of the handle buttons
C ^T
Do Compliance
Maintains compliance until ^F
C ^U
Power Down
Powers down unit
C ^W
Stop Compliance
Turn off compliance