No�
Fault
Can be recognized by
Possible cause
Remedial measures
5
Connector wobbles despite
a correct connection to the
test piece
Incorrect connection
geometry at test piece
Wear on geometry of the
test piece’s connection
parts
Replace test piece
Wear on the connector’s
clamping jaws
Return the connector to WEH
for maintenance
6
Clamping jaws deformed
The connector can not
be plugged into the test
piece
Excessive mechanical
stress due to inappropri-
ate handling
Return the connector to WEH
for maintenance
7
The clamping lever can no longer be actuated
Connector is not correct-
ly connected
Repeat the connecting procedure
Damaged components
Return the connector to WEH for
maintenance
Inadequate lubrication
Lubricate the connector, see
Chapter 10. Lubrication.
If you should encounter any other problems, please contact WEH or your responsible distributor.