12
9. The instrument will take 1 silver reading and 1 copper reading and
store the reading as calibration data. When the reading is captured,
press
Next
.
10. Remove tube from Solution C bottle and reattach tube to peristaltic
pump. Restore the water supply and click
Next
.
Calibration is now complete. Calibration data is saved after all steps
are complete. If calibration is canceled at any point in the calibration
process, the last successful calibration data remains stored. The CSM
returns to normal operation once calibration is complete.
8.0 Accessories and Replacement
Parts List
Catalog Number
Accessory Description
25018S
Replacement Cuvette
28062
CSM Instrument
28141S
Pump Head Replacement Kit
28159S
Calibration Kit
28190S
Pressure Regulator Assembly
28162S
Tubing Replacement Kit
28163S
Reagent Kit
28164S
Air Pump Kit
28167S
Reagent Pump Kit
100184
Operation Manual, CSM
100003S
Solenoid Valve Kit
110154S
Tethered Light Shield Replacement Kit
To order any accessory or replacement parts, please contact the HF
scientific Customer Service Department.
HF scientific
16260 Airport Park Drive, Suite 140
Fort Myers, FL 33913
Phone: 239-337-2116
Toll Free: 888-203-7248
Fax: 239-454-0694
Email: [email protected]
Website: www.hfscientific.com
4. Disconnect the upper inlet tubing hose from the peristaltic pump and
attach a single piece of black tubing included in the calibration kit to
the upper inlet on the peristaltic pump. Insert the opposite end of the
tubing into Solution A bottle (Low Range Copper/Silver). Press
Next
.
5. During this time, the software will take 1 silver reading and 1 copper
reading and store the readings as calibration data. When the reading
is captured, press
Next
.
6. Remove tube from Solution A bottle, rinse with Solution D (DI water).
Insert tube into Solution B bottle (Mid Range Silver). Press
Next
.
7. The instrument will take 1 silver reading and store the reading as
calibration data. When the reading is captured, press
Next
.
8. Remove tube from Solution B bottle, rinse with Solution D. Insert
tube into Solution C bottle (High Range Copper/Silver). Press
Next
.