background image

 

8

5.6 Setting the Reading Mode

The CSM has 2 different modes of taking readings, The first is Auto 

Mode. In Auto Mode, measurements are taken at a preset time interval 

and displayed on the Home screen. A countdown clock showing the 

time until the next reading is displayed in the status bar on the bottom 

of the home screen. 
The second mode is Manual Mode. In Manual Mode, the date/time 

button on the home screen is used to take an on-demand reading. The 

date/time button displays the date and time of the last reading while 

the measurement data is shown to the right. The status bar shows 

GOOD, rather than a countdown clock. 
To set the Reading Mode:
1.  Press 

Measure Ag, Cu, Ag/Cu

 on the Configuration Menu screen.

2.  Press 

Auto

 or 

Manual

3.  Press which parameter you wish to monitor: Silver, Copper or both. 

Then press 

Save

.

4.  Once the desired Reading Mode has been saved, press the 

 

button to return to the home screen or press 

Cancel

 to return to the 

previous menu.

5.7 Adjusting the Line Flush Duration

A line flush is necessary to wash the cuvette between measurements. 

The line flush setting is used to set the amount of time in seconds to 

wash the cuvette. This time will be applied to the wash step in the 

reading sequence, discussed later in this manual. The default duration 

is 5 seconds, which is also the minimum. The maximum line flush 

duration is 60 seconds. The suggested line flush duration is 1 second 

for every foot of sample line tubing (assuming inlet pressure of 4-60psi)
To adjust the line flush duration:
1.  Press 

Line Flush Setting

 on the Configuration Menu screen.

2.  Using the up and down arrows, set the desired duration for the line 

flush in seconds. 

3.  Once the desired Line Flush Duration has been saved, press the 

 

button to return to the home screen or press 

Cancel

 to return to the 

previous menu.

5.8 Security Setting

The CSM has a security setting that when enabled will only allow 

authorized users with a valid numeric password to configure the 

instrument. The user must enter a 6 digit PIN to gain admin access. 

There is a Master PIN that is hard-coded into the software and it 

cannot be changed. If a user forgets their personal pin, HF can provide 

the Master PIN and the instrument will prompt the user to choose a 

new PIN. Contact HF scientific Technical Service Department for the 

Master PIN.
To enable a Security PIN:
1.  Press 

Security

 on the Configuration Menu screen.

2.  If this is the first time entering the Security PIN, enter the Master 

PIN provided by HF scientific. If you already have a personal PIN 

assigned, enter that PIN. Press 

ENTER

.

3.  Press Enable Admin or Change PIN. 

4.  If selecting to 

Enable Admin

, enter the security PIN again to 

enable. If selecting 

Change PIN

, enter the new pin. You will be 

prompted to enter the new pin a second time. Then press 

Enter

.

5.  Once the security PIN or setting has been saved, press the 

 

button to return to the home screen or press 

Cancel

 to return to the 

previous menu.

Summary of Contents for 28062

Page 1: ...ilver Monitor IS HF CSM 100184 WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage or damage to the equipment Keep this Manual for future reference Rev 1 ...

Page 2: ...ing the Units of Measurement 7 5 3 Setting the Date Time 7 5 4 Adjusting the Brightness 7 5 5 Setting the Time Schedule 7 5 6 Setting the Reading Mode 8 5 7 Adjusting the Line Flush Duration 8 5 8 Security Setting 8 5 9 Firmware Update 9 5 10 Copying FRAM data to SD Card 9 5 11 Factory Reset 10 5 12 About Device 10 6 0 System Start Up 10 6 1 Prepare Reagents 10 6 2 Priming the CSM 10 6 3 Taking Re...

Page 3: ...nt This pictorial alerts you to the need to read the manual This pictorial alerts you to electricity electrocution and shock hazards 1 0 Understanding Safety Information This manual contains safety and use instructions that must be followed during the installation commissioning operation care and maintenance and service of the CSM All responsible personnel must read this manual prior to working wi...

Page 4: ...o 23 hours 59 minutes Path Length 10mm Display 3 5 Color graphic touch screen Analog Output Powered 4 20mA 1000 Ω drive isolated Maximum Water Pressure 124 kPa 18 psi see Plumbing Connections section Flow Rate 185 ml min to 225 ml min 044 to 059 Gal min Operating Temperature 1 C 50 C 34 F 122 F Max Sample Water Intake and Drainage Flow 225 ml minute Wetted Materials PBT Polypropylene 316 Stainless...

Page 5: ...be removed for easier connections Please ensure this terminal set is properly tightened using the two side screws 3 2 Plumbing Connections All plumbing connections are made with user supplied OD tubing using the push to connect fittings attached to the CSM All water tubing used to connect to the CSM is supplied by the installer Sample water should be particulate free The sample water supply does n...

Page 6: ...electrical panel and tighten Securing the Electrical Enclosure Once all electrical connections are made the electrical panel must be secured by closing the compartment and tightening the captive Phillips screw 4 0 Commissioning and Start Up 1 Once all water and electrical connections have been made turn on power to the CSM 2 The CSM will run through a startup process that takes less than 10 second...

Page 7: ...ve 4 Once the date and time have been saved press the button to return to the home screen or press Cancel to return to the previous menu 5 4 Adjusting the Brightness The screen brightness can be adjusted to provide easier readability of the display in low light or no light conditions There are 10 brightness levels available To adjust the screen brightness 1 Press Brightness on the Configuration Me...

Page 8: ... suggested line flush duration is 1 second for every foot of sample line tubing assuming inlet pressure of 4 60psi To adjust the line flush duration 1 Press Line Flush Setting on the Configuration Menu screen 2 Using the up and down arrows set the desired duration for the line flush in seconds 3 Once the desired Line Flush Duration has been saved press the button to return to the home screen or pr...

Page 9: ...ate is complete a restart of the instrument is required Do not remove the SD card or turn power off while the software update is in progress Doing so can cause an unrecoverable failure NOTICE 5 10 Copying FRAM data to SD Card The following information from the CSM can be transferred to an SD card in CSV file format Readings Error and warning logs with date and time stamps All user settings Calibra...

Page 10: ...he prime sub menu 3 Press the desired priming option If this is the first time starting up the instrument start with Prime Water and repeat this step for all priming options in this menu Reagent Bags Hanging on Brackets P N 110097 Assembled in the USA Reagent Kit Copper Indicator 16260 Airport Park Drive Suite 140 Fort Myers FL 33913 L 74125 Rev E 11 2020 WARNING Harmful if swallowed Causes skin i...

Page 11: ...per Silver 500 ml Solution C in Cal Procedure DI Water 500 ml Solution D in the Cal Procedure To recalibrate the CSM follow the prompts on the screen 1 From the screen press Service to enter the Service Menu 2 Press CALIBRATE to begin the Calibration sequence Ensure you have the calibration kit ready then press NEXT 3 Turn off the water supply Press Next There is a removable light shield attached ...

Page 12: ...placement Kit To order any accessory or replacement parts please contact the HF scientific Customer Service Department HF scientific 16260 Airport Park Drive Suite 140 Fort Myers FL 33913 Phone 239 337 2116 Toll Free 888 203 7248 Fax 239 454 0694 Email hf info wattswater com Website www hfscientific com 4 Disconnect the upper inlet tubing hose from the peristaltic pump and attach a single piece of...

Page 13: ...gent tubing skip to step 16 12 Insert short piece of new tubing into outlet fitting on solenoid valve Insert other end into inlet fitting into optical injector 13 Secure solenoid valve to bracket with 2 screws removed in step 6 14 Insert end of new tubing into inlet of the solenoid valve 15 Reuse Velcro strap to secure the reagent tubing check valves to the top of the pump bracket 16 Guide tubing ...

Page 14: ...the base 9 4 Replacing the Pressure Regulator Assembly Kit 28190S To replace the pressure regulator assembly on the CSM instrument 1 Turn off water to the CSM instrument 2 Loosen the screws on the bracket connecting the pressure regulator to the back panel 3 Lift bracket and attached components from back panel 4 Loosen the water inlet tube orange from the pressure regulator 5 Remove the panel nut ...

Page 15: ...pump to the optical board and then remove air pump 11 Insert new air pump Be sure the arrow on the new air pump is facing the correct direction 12 Install 2 screws from underneath air pump bracket 13 Attach air pump to optical board using the included zip tie 14 Screw in black fitting into air pump outlet 15 Reconnect the black and pink air pump wiring connection located in the electrical panel 16...

Page 16: ...IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE The remedy described in the first paragraph of this warranty shall constitute the sole and exclusive remedy for breach of warranty and the Company shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which ...

Reviews: