Table 13 - Pump Control
ADI
Name
Access
Type
Description
2
Set pump
speed
(deciRPM)
Write
UInt16
Speed is set in Deci RPM. Max speed depends on head
type. See Pump Head enumeration table
3
Set pump
speed limit
(deciRPM)
Write
UInt16
Speed is set in Deci RPM. Max speed depends on head
type. See Pump Head enumeration table
4
Set failsafe
speed
(deciRPM)
Write
UInt16
If the failsafe is enabled the pump will run continuously
at the selected speed in the event of a communications
loss.
101
Control
bitfield
Write
UInt16
Bit 0
Set fail safe enable, Enabled the failsafe
speed. If disabled the pump will stop in the
event of a communications loss. If enabled
the pump will run at the speed set in the
“SetFailsafeSpeed” parameter
Bit 1
Set pump direction to anti-clockwise, If set
the pump will run anti-clockwise. Pump
defaults to clockwise rotation
Bit 2
Start pump, Set to 1 (true) to allow the
pump to run. 0 will stop the pump. Note
that pump enable needs to be set
Bit 3
Enable pump, Need to set to 1 to allow
pump to run. Setting to 0 will stop the
pump and not allow the pump to run.
Bit 4
Reset pump run hours to zero, Resets the
run hours accumulator
Bit 5
Pause flow totaliser, Set to 1 to pause the
internal Total volume pumped parameter.
Setting to 0 will un-pause the parameter
Bit 6
Reset flow totaliser to zero, Set to 1 to reset
the Total volume pumped to 0. Set to 0 to
allow the Total volume pumped to
accumulate
Bit 7
Reset revolution count to zero, Set to 1 to
reset the Pump head revolution count to 0.
Set to 0 to allow the Pump head revolution
count to increment.
88
m-530pn-en-09 02-02-22