Waters CORTECS Care And Use Manual Download Page 9

9

CORTECS and CORTECS Premier 2.7 

µ

m Columns

[ CARE AND USE MANUAL ]

highest quality solvents possible. HPLC-grade water 
is not UPLC-grade water.

 

Consider sample preparation (e.g., solid-phase 
extraction, filtration, etc.).

5.  Avoid (where possible):

 

100% aqueous mobile phases (if possible)

 

HPLC-grade bottled water

 

“Topping off” or adding “new” mobile phase 
to “old” mobile phase

 

Old aqueous mobile phases. Remember to rinse bottles 
thoroughly and prepare fresh every 24 to 48 hours

 

Using phosphate salt buffer in combination with high 
ACN concentrations (e.g., >70%) due to precipitation.

6.  Don’t assume a “bad” column is the culprit when high 

backpressure or split peaks are observed. Investigate 
cause of column failure by checking:

 

Backpressure

 

Mobile phase(s), bacteria, precipitation, 
and/or samples

 

Peak splitting

 

Sample quality

 

Injection solvent strength

7.  To reduce the chances of mobile-phase contamination or 

degradation, only prepare what you need for analysis or 
store excess bulk quantities in a refrigerated environment.

8.  Mobile phase-related questions to ask:

 

Am I using 100% aqueous mobile phases? 
Am I able to add a small amount of organic 
modifier to my mobile phase A?

 

Do I filter my aqueous mobile phases through 
0.2 µm filters?

 

How old is my mobile phase? Do I label the bottle 
with the preparation date?                                                                                                                                      

 

Do I “top off” or do I prepare fresh mobile phases 
every 24–48 hours?

 

What is the quality of my water? Has the quality 
recently changed? How is my water purification 
system working? When was it last serviced?

 

Am I working with pH 7 phosphate buffer (which 
is VERY susceptible to bacterial growth)?

9.  Sample-related questions to ask:

 

If I inject neat standards prepared in mobile 
phase do I observe these problems?

 

If I prepare my standards in water and prepare them 
like samples  (e.g., SPE, filtration, etc.) do I still observe 
these problems? 

 

Has the quality of my samples changed over time?

b. Getting Started with CORTECS HILIC Columns

1.  Because CORTECS HILIC columns do not possess a 

bonded phase, the pH operating range is 1–5, and they can  
be operated at temperatures up to 45 °C.

2.  As with any LC column, operating at the extremes of pH, 

pressures, and temperatures will result in decreased 
column lifetime.

Column Equilibration

1.  When the column is first received, flush in 50% acetonitrile: 

50% water with 10 mM final buffer concentration for 
50 column volumes. 

2.  Equilibrate with 20-column volumes of initial mobile-phase 

conditions before making first injection. 

3.  If gradient conditions are used, equilibrate with 

8–10 column volumes between injections.

4.  Failure to appropriately equilibrate the column could 

result in drifting retention times.

Mobile-phase Considerations

1.  Always maintain at least 3% polar solvent in the mobile 

phase or gradient (e.g., 3% aqueous/3% methanol or 
2% aqueous/1% methanol, etc.). This ensures that the 
CORTECS particle is always hydrated.

2.  Maintain at least 40% organic solvent (e.g., acetonitrile) 

in your mobile phase or gradient.

3.  Avoid phosphate salt buffers to avoid precipitation in 

HILIC mobile phases. Phosphoric acid is okay.

4.  Buffers such as ammonium formate or ammonium acetate 

will produce more reproducible results than additives such 
as formic acid or acetic acid. If an additive (e.g., formic 
acid, etc.) must be used instead of a buffer, use 0.2% (v:v) 
instead of 0.1%.

5.  For best peak shape, maintain a buffer concentration 

of 10 mM in your mobile phase/gradient at all times. 

Summary of Contents for CORTECS

Page 1: ...CORTECS packing material is tested chromatographically with acidic basic and neutral analytes and the results are held to narrow specification ranges to assure excellent reproducible performance Ever...

Page 2: ...lumn outlet to the detector This prevents entry of air into the detection system 4 Gradually increase the flow rate as described in Step 2 5 Once a steady backpressure and baseline have been achieved...

Page 3: ...guard holder and guard cartridge can be easily attached to the inlet of the column For more information on VanGuard products visit www waters com vanguard g VanGuard FIT Cartridges The VanGuard FIT C...

Page 4: ...ating temperatures are recommended for longer column lifetime Working at high temperatures e g 45 C may also result in shorter column lifetimes III COLUMN USE To ensure the continued high performance...

Page 5: ...at 10 mM or less Ammonia NH4 OH Ammonium bicarbonate 9 2 10 3 HCO3 9 2 NH4 8 2 10 2 8 2 11 3 Volatile Volatile Yes Yes Used in the 5 10 mM range for MS work keep source 150 C Adjust pH with ammonium...

Page 6: ...rface charge modification End cap style Carbon load Ligand density pH limits Temp limits C18 Trifunctional C18 None Proprietary 6 6 2 7 mol m2 2 8 45 C C18 Trifunctional C18 Proprietary 5 7 2 4 mol m2...

Page 7: ...rmate containing mobile phase V eCORD INTELLIGENT CHIP TECHNOLOGY May not be available for all column configurations a Introduction The eCord Intelligent Chip provides the history of a column s perfor...

Page 8: ...ple set maximum pressure and temperature in the sample set and if the column met basic system suitability requirements Up to 50 sample sets can be stored on the eCord chip VI ADDITIONAL INFORMATION a...

Page 9: ...related questions to ask If I inject neat standards prepared in mobile phase do I observe these problems If I prepare my standards in water and prepare them like samples e g SPE filtration etc do I s...

Page 10: ...es are in the 0 4 to 0 8 mL min range 3 As compared to Atlantis HILIC Silica HPLC Columns the CORTECS HILIC columns are approximately 20 less retentive for gradient analysis and 35 65 less retentivefo...

Reviews: