Waters CORTECS Care And Use Manual Download Page 8

8

CORTECS and CORTECS Premier 2.7 

µ

m Columns

[ CARE AND USE MANUAL ]

c. Manufacturing Information

Figure 6. The eCord chip provides 

the user with QC test conditions 

and results on the column run by 

the manufacturer. The information 

includes mobile phases, running 

conditions, and analytes used 

to test the columns. In addition, 

the QC results and acceptance 

is placed onto the column.

Figure 5. The eCord chip provides the user with 

an overview of the bulk material QC test results.

Figure 7. An example of column use information provided by the eCord chip. 

d. Column Use Information

The eCord chip provides the customer with column use 
data. The top of the screen identifies the column including 
chemistry type, column dimensions and serial number. The 
overall column usage information includes the total number 
of samples, total number of injections, total sample sets, date 
of first injection, date of last injection, maximum pressure, and 
temperature. The information also details the column history 
by sample set including date started, sample set name, user 
name, system name, number of injections in the sample set, 
number of samples in the sample set, maximum pressure, and 
temperature in the sample set and if the column met basic 
system suitability requirements. Up to 50 sample sets can be 
stored on the eCord chip.

VI. ADDITIONAL INFORMATION

a. Tips for Maximizing CORTECS Column Lifetimes

1.  To maximize CORTECS Column lifetime, pay close 

attention to:

 

Water quality (including water purification system)

 

Solvent quality

 

Mobile-phase preparation, storage, and age

 

Sample, buffer, and mobile-phase solubilities

 

Sample quality and preparation

2.  It is recommended to monitor your systems health by 

incorporating a quality control (QC) reference material 
into your work flow. (Reference 720004535en)

3.  When problems arise, often only one improper practice 

must be changed.

4.  Always remember to:

 

Use in-line filter unit or, preferably, 
a VanGuard Cartridge.

 

Discourage bacterial growth by minimizing the use 
of 100% aqueous mobile phases where possible.

 

Change aqueous mobile phase every 24–48 hours 
(if 100% aqueous mobile phase use is required).

 

Discard old 100% aqueous mobile phases every 
24–48 hours to discourage bacterial growth.

 

Add 5–10% organic modifier to mobile 
phase A and adjust gradient profile.

 

Filter aqueous portions of mobile phase 
through 0.2 µm filter.

 

Maintain your water purification system so that 
it is in good working order.

 

Only use ultra pure water (18 megohm-cm) water and 

Summary of Contents for CORTECS

Page 1: ...CORTECS packing material is tested chromatographically with acidic basic and neutral analytes and the results are held to narrow specification ranges to assure excellent reproducible performance Ever...

Page 2: ...lumn outlet to the detector This prevents entry of air into the detection system 4 Gradually increase the flow rate as described in Step 2 5 Once a steady backpressure and baseline have been achieved...

Page 3: ...guard holder and guard cartridge can be easily attached to the inlet of the column For more information on VanGuard products visit www waters com vanguard g VanGuard FIT Cartridges The VanGuard FIT C...

Page 4: ...ating temperatures are recommended for longer column lifetime Working at high temperatures e g 45 C may also result in shorter column lifetimes III COLUMN USE To ensure the continued high performance...

Page 5: ...at 10 mM or less Ammonia NH4 OH Ammonium bicarbonate 9 2 10 3 HCO3 9 2 NH4 8 2 10 2 8 2 11 3 Volatile Volatile Yes Yes Used in the 5 10 mM range for MS work keep source 150 C Adjust pH with ammonium...

Page 6: ...rface charge modification End cap style Carbon load Ligand density pH limits Temp limits C18 Trifunctional C18 None Proprietary 6 6 2 7 mol m2 2 8 45 C C18 Trifunctional C18 Proprietary 5 7 2 4 mol m2...

Page 7: ...rmate containing mobile phase V eCORD INTELLIGENT CHIP TECHNOLOGY May not be available for all column configurations a Introduction The eCord Intelligent Chip provides the history of a column s perfor...

Page 8: ...ple set maximum pressure and temperature in the sample set and if the column met basic system suitability requirements Up to 50 sample sets can be stored on the eCord chip VI ADDITIONAL INFORMATION a...

Page 9: ...related questions to ask If I inject neat standards prepared in mobile phase do I observe these problems If I prepare my standards in water and prepare them like samples e g SPE filtration etc do I s...

Page 10: ...es are in the 0 4 to 0 8 mL min range 3 As compared to Atlantis HILIC Silica HPLC Columns the CORTECS HILIC columns are approximately 20 less retentive for gradient analysis and 35 65 less retentivefo...

Reviews: