Waters CORTECS Care And Use Manual Download Page 7

7

CORTECS and CORTECS Premier 2.7 

µ

m Columns

[ CARE AND USE MANUAL ]

b. Storage 

For periods longer than four days at room temperature, store 
reversed-phase columns in 100% acetonitrile. For elevated 
temperature applications, store immediately after use in 100% 
acetonitrile for the best column lifetime. Do not store columns 
in buffered eluents. If the mobile phase contained a buffer salt, 
flush reversed-phase CORTECS Columns with 10-column 
volumes of HPLC grade water (see Table 1 for common column 
volumes) and replace with 100% acetonitrile for storage. Failure 
to perform this intermediate step could result in precipitation 
of the buffer salt in the column when 100% acetonitrile is 
introduced. Completely seal column to avoid evaporation and 
drying out of the bed. 

For periods longer than four days, store CORTECS Columns 
in 95:5 acetonitrile:water. Do not store in buffered solvent. If 
the mobile phase contained a buffered salt, flush the column 
with 10-column volumes of 95:5 acetonitrile:water (see Table 1 
for common column volumes).

Note: If a column has been run with a mobile phase that contains 
formate (e.g., ammonium formate, formic acid, etc.) and is then 
flushed with 100% acetonitrile, slightly longer equilibration times 
may be necessary when the column is re-installed and run again 
with a formate-containing mobile phase.

V. eCORD INTELLIGENT CHIP TECHNOLOGY 

(May not be available for all column configurations)

a. Introduction

The eCord Intelligent Chip provides the history of a column’s 
performance throughout its lifetime. The eCord will be 
permanently attached to the column to assure that the 
column’s performance history is maintained in the event that 
the column is moved from one instrument to another.  

Figure 3. eCord Intelligent Chip.

eCord Intelligent Chip

Figure 4. eCord inserted into side of column heater. 

eCord inserted into 

side of column heater

At the time of manufacture, tracking, and quality control 
information will be downloaded to the eCord. Storing this 
information on the chip will eliminate the need for a paper 
Certificate of Analysis. Once the user installs the column, 
the software will automatically download key parameters 
into a column history file stored on the chip. In this manual, 
we explain how the eCord will provide a solution for easily 
tracking the history of the columns, reduce the frustration of 
paperwork trails, and give customers the reassurance that a 
well-performing column is installed onto their instruments.

b. Installation

Install the column into the column heater. Plug the eCord into 
the side of the column heater. Once the eCord is inserted into 
the column heater the identification and overall column usage 
information will be available in Empower™ and MassLynx™ 
Software allowing the user to access column information on  
their desktop.

Summary of Contents for CORTECS

Page 1: ...CORTECS packing material is tested chromatographically with acidic basic and neutral analytes and the results are held to narrow specification ranges to assure excellent reproducible performance Ever...

Page 2: ...lumn outlet to the detector This prevents entry of air into the detection system 4 Gradually increase the flow rate as described in Step 2 5 Once a steady backpressure and baseline have been achieved...

Page 3: ...guard holder and guard cartridge can be easily attached to the inlet of the column For more information on VanGuard products visit www waters com vanguard g VanGuard FIT Cartridges The VanGuard FIT C...

Page 4: ...ating temperatures are recommended for longer column lifetime Working at high temperatures e g 45 C may also result in shorter column lifetimes III COLUMN USE To ensure the continued high performance...

Page 5: ...at 10 mM or less Ammonia NH4 OH Ammonium bicarbonate 9 2 10 3 HCO3 9 2 NH4 8 2 10 2 8 2 11 3 Volatile Volatile Yes Yes Used in the 5 10 mM range for MS work keep source 150 C Adjust pH with ammonium...

Page 6: ...rface charge modification End cap style Carbon load Ligand density pH limits Temp limits C18 Trifunctional C18 None Proprietary 6 6 2 7 mol m2 2 8 45 C C18 Trifunctional C18 Proprietary 5 7 2 4 mol m2...

Page 7: ...rmate containing mobile phase V eCORD INTELLIGENT CHIP TECHNOLOGY May not be available for all column configurations a Introduction The eCord Intelligent Chip provides the history of a column s perfor...

Page 8: ...ple set maximum pressure and temperature in the sample set and if the column met basic system suitability requirements Up to 50 sample sets can be stored on the eCord chip VI ADDITIONAL INFORMATION a...

Page 9: ...related questions to ask If I inject neat standards prepared in mobile phase do I observe these problems If I prepare my standards in water and prepare them like samples e g SPE filtration etc do I s...

Page 10: ...es are in the 0 4 to 0 8 mL min range 3 As compared to Atlantis HILIC Silica HPLC Columns the CORTECS HILIC columns are approximately 20 less retentive for gradient analysis and 35 65 less retentivefo...

Reviews: