Waters CORTECS Care And Use Manual Download Page 4

4

CORTECS and CORTECS Premier 2.7 

µ

m Columns

[ CARE AND USE MANUAL ]

b. pH Range

The recommended operating pH ranges for CORTECS 
columns can be found in table 3. A listing of commonly used 
buffers and additives is given in Table 2. Additionally, the 
column lifetime will vary depending upon the operating 
temperature, the type, and concentration of buffer used. For 
example, the use of phosphate buffer at 
pH 8 in combination with elevated temperatures will lead 
to shorter column lifetimes.

c. Solvents 

To maintain maximum column performance, use high quality 
chromatography grade solvents. Filter all aqueous buffers prior 
to use through a 0.2 µm filter. Pall Gelman Laboratory Acrodisc® 
filters are recommended. Solvents containing suspended 
particulate materials will generally clog the outside surface of 
the inlet distribution frit of the column. This will result in higher 
operating pressure and poorer performance. See Section VI 
for more information.

d. Pressure 

1. The 2.1 mm I.D. CORTECS and CORTECS Premier 2.5 µm
  Columns can tolerate operating pressures up to 18,000  
  psi (1241 bar or 124 MPa).
2. The 3.0 mm I.D. CORTECS Columns can tolerate   
operating pressures up to 18,000 psi 
  (1241 bar or 124 MPa).
3. The 4.6 mm I.D. CORTECS and CORTECS 2.5 µm
  Columns can tolerate pressures up to 10,000 psi
  (700 bar or 70 MPa).

e. Temperature 

Temperatures between 20–45 ˚C are recommended 
for operating CORTECS Columns in order to enhance 
selectivity, lower solvent viscosity, and increase mass 
transfer rates. When operating at high pH, lower operating 
temperatures are recommended for longer column lifetime. 
Working at high temperatures (e.g., >45 °C) may also result in 
shorter column lifetimes. 

III. COLUMN USE

To ensure the continued high performance of CORTECS 
Columns, follow these guidelines: 

a. Sample Preparation 

1.  Sample impurities often contribute to column 

contamination. One option to avoid this is to use Waters 
Oasis™ Solid-phase Extraction Cartridges/columns or 
Sep-Pak™ Cartridges of the appropriate chemistry to 
clean up the sample before analysis. For more information, 
visit 

www.waters.com/sampleprep

.

2.  It is preferable to prepare the sample in the operating 

mobile phase or a mobile phase that is weaker than the 
mobile phase for the best peak shape and sensitivity. 
Acetone should not be used as a sample solvent/diluent 
unless a Hexane Tetrahydrofuran Compatibility Kit has 
been installed.

3.  If the sample is not dissolved in the mobile phase, ensure 

that the sample, solvent, and mobile phases are miscible 
in order to avoid sample and/or buffer precipitation.

4.  Filter sample with 0.2 µm membranes to remove particulates. 

If the sample is dissolved in a solvent that contains an organic 
modifier (e.g., acetonitrile, methanol, etc.) ensure that the 
membrane material does not dissolve in the solvent. Contact 
the membrane manufacturer with solvent compatibility 
questions. Alternatively, centrifugation for 20 minutes at 
8000 rpm, followed by the transfer of the supernatant liquid 
to an appropriate vial, could be considered.

5.  For Hydrophilic Interaction Chromatography (HILIC) 

separations, the samples must be prepared in a high 
percentage of organic solvents (e.g., 95% acetonitrile). 
See “Getting Started with CORTECS HILIC Columns” 
section for additional information.

Summary of Contents for CORTECS

Page 1: ...CORTECS packing material is tested chromatographically with acidic basic and neutral analytes and the results are held to narrow specification ranges to assure excellent reproducible performance Ever...

Page 2: ...lumn outlet to the detector This prevents entry of air into the detection system 4 Gradually increase the flow rate as described in Step 2 5 Once a steady backpressure and baseline have been achieved...

Page 3: ...guard holder and guard cartridge can be easily attached to the inlet of the column For more information on VanGuard products visit www waters com vanguard g VanGuard FIT Cartridges The VanGuard FIT C...

Page 4: ...ating temperatures are recommended for longer column lifetime Working at high temperatures e g 45 C may also result in shorter column lifetimes III COLUMN USE To ensure the continued high performance...

Page 5: ...at 10 mM or less Ammonia NH4 OH Ammonium bicarbonate 9 2 10 3 HCO3 9 2 NH4 8 2 10 2 8 2 11 3 Volatile Volatile Yes Yes Used in the 5 10 mM range for MS work keep source 150 C Adjust pH with ammonium...

Page 6: ...rface charge modification End cap style Carbon load Ligand density pH limits Temp limits C18 Trifunctional C18 None Proprietary 6 6 2 7 mol m2 2 8 45 C C18 Trifunctional C18 Proprietary 5 7 2 4 mol m2...

Page 7: ...rmate containing mobile phase V eCORD INTELLIGENT CHIP TECHNOLOGY May not be available for all column configurations a Introduction The eCord Intelligent Chip provides the history of a column s perfor...

Page 8: ...ple set maximum pressure and temperature in the sample set and if the column met basic system suitability requirements Up to 50 sample sets can be stored on the eCord chip VI ADDITIONAL INFORMATION a...

Page 9: ...related questions to ask If I inject neat standards prepared in mobile phase do I observe these problems If I prepare my standards in water and prepare them like samples e g SPE filtration etc do I s...

Page 10: ...es are in the 0 4 to 0 8 mL min range 3 As compared to Atlantis HILIC Silica HPLC Columns the CORTECS HILIC columns are approximately 20 less retentive for gradient analysis and 35 65 less retentivefo...

Reviews: