CORTECS and CORTECS Premier 2.7 

µ

m Columns

[ CARE AND USE MANUAL ]

Table 2. Buffer Recommendations for Using CORTECS Columns

Additive/Buffer

pK

a

Buffer 

Range  

Volatility 

(±1 pH Unit)

Used for 

Mass 

Spec

Comments

TFA

0.3

Volatile

Yes

Ion pair additive, can suppress MS signal, 

used in the 0.02–0.1% range.

Acetic acid

4.76

Volatile

Yes

Maximum buffering obtained when used with 

ammonium acetate salt. Used in 0.1–1.0% range.

Formic acid

3.75

Volatile

Yes

Maximum buffering obtained when used with 

ammonium formate salt. Used in 0.1–1.0% range.

Acetate (NH

4

CH

3

CO

2

)

4.76

3.76–5.76

Volatile

Yes

Used in the 1–10 mM range. 

Note: Sodium or potassium salts are not volatile.

Formate (NH

4

HCO

2

)

3.75

2.75–4.75

Volatile

Yes

Used in the 1–10 mM range. 

Note: Sodium or potassium salts are not volatile.

Phosphate 1

2.15

1.15–3.15 Non-volatile

No

Traditional low pH buffer, good UV transparency.

Phosphate 2

7.2

6.20–8.20 Non-volatile

No

Above pH 7, reduce temperature/concentration 

and use a guard column to maximize lifetime.

Phosphate 3

12.3

11.3–13.3 Non-volatile

No

Above pH 7, reduce temperature/concentration 

and use a guard column to maximize lifetime.

4-Methylmorpholine

~8.4

7.4–9.4

Volatile

Yes

Generally used at 10 mM or less.

Ammonia (NH

4

OH)

Ammonium 

bicarbonate

9.2

10.3 (HCO

3

-

)

9.2 (NH

4

+

)

8.2–10.2

8.2–11.3

Volatile

Volatile

Yes

Yes

Used in the 5–10 mM range (for MS work keep 

source >150 ˚C). Adjust pH with ammonium 

hydroxide or acetic acid. Good buffering 

capacity at pH 10. 

Note: Use ammonium bicarbonate (NH

4

HCO

3

), 

not ammonium carbonate ((NH

4

)

2

CO

3

).

Ammonium (Acetate)

9.2

8.2–10.2

Volatile

Yes

Used in the 1–10 mM range.

Ammonium (Formate)

9.2

8.2–10.2

Volatile

Yes

Used in the 1–10 mM range.

Borate

9.2

8.2–10.2 Non-volatile

No

Reduce temperature/concentration and use a 

guard column to maximize lifetime.

CAPSO

9.7

8.7–10.7

Non-volatile

No

Zwitterionic buffer, compatible with acetonitrile, 

used in the 1–10 mM range. Low odor.

Glycine

2.4, 9.8

8.8–10.8 Non-volatile

No

Zwitterionic buffer, can give longer 

lifetimes than borate buffer.

1-Methylpiperidine

10.2

9.3–11.3

Volatile

Yes

Used in the 1–10 mM range.

CAPS

10.4

9.5–11.5

Non-volatile

No

Zwitterionic buffer, compatible with acetonitrile, 

used in the 1–10 mM range. Low odor. 

Triethylamine 

(as acetate salt)

10.7

9.7–11.7

Volatile

Yes

Used in the 0.1-1.0% range. Volatile only when 

titrated with acetic acid (not hydrochloric 

or phosphoric). Used as ion-pair for DNA 

analysis at pH 7–9.

Pyrrolidine

11.3

10.3–12.3

Volatile

Yes

Mild buffer, gives long lifetime.

 

Note: Working at the extremes of pH, temperature, and/or pressure will result in shorter column lifetimes.

Summary of Contents for CORTECS

Page 1: ...CORTECS packing material is tested chromatographically with acidic basic and neutral analytes and the results are held to narrow specification ranges to assure excellent reproducible performance Ever...

Page 2: ...lumn outlet to the detector This prevents entry of air into the detection system 4 Gradually increase the flow rate as described in Step 2 5 Once a steady backpressure and baseline have been achieved...

Page 3: ...guard holder and guard cartridge can be easily attached to the inlet of the column For more information on VanGuard products visit www waters com vanguard g VanGuard FIT Cartridges The VanGuard FIT C...

Page 4: ...ating temperatures are recommended for longer column lifetime Working at high temperatures e g 45 C may also result in shorter column lifetimes III COLUMN USE To ensure the continued high performance...

Page 5: ...at 10 mM or less Ammonia NH4 OH Ammonium bicarbonate 9 2 10 3 HCO3 9 2 NH4 8 2 10 2 8 2 11 3 Volatile Volatile Yes Yes Used in the 5 10 mM range for MS work keep source 150 C Adjust pH with ammonium...

Page 6: ...rface charge modification End cap style Carbon load Ligand density pH limits Temp limits C18 Trifunctional C18 None Proprietary 6 6 2 7 mol m2 2 8 45 C C18 Trifunctional C18 Proprietary 5 7 2 4 mol m2...

Page 7: ...rmate containing mobile phase V eCORD INTELLIGENT CHIP TECHNOLOGY May not be available for all column configurations a Introduction The eCord Intelligent Chip provides the history of a column s perfor...

Page 8: ...ple set maximum pressure and temperature in the sample set and if the column met basic system suitability requirements Up to 50 sample sets can be stored on the eCord chip VI ADDITIONAL INFORMATION a...

Page 9: ...related questions to ask If I inject neat standards prepared in mobile phase do I observe these problems If I prepare my standards in water and prepare them like samples e g SPE filtration etc do I s...

Page 10: ...es are in the 0 4 to 0 8 mL min range 3 As compared to Atlantis HILIC Silica HPLC Columns the CORTECS HILIC columns are approximately 20 less retentive for gradient analysis and 35 65 less retentivefo...

Reviews: